![Delta OHM DO 9847K Скачать руководство пользователя страница 60](http://html1.mh-extra.com/html/delta-ohm/do-9847k/do-9847k_manual_2478269060.webp)
- -
60
Technical Specifications
AP472 S1
AP472 S2
AP472 S4…
L LT H HT
Type of Measurements
Air velocity,
calculated flow,
air temperature
Air velocity,
calculated flow
Air v
elo
city,
calculated fl
ow
Air v
elo
city,
calculated fl
ow,
air tempe
rature
Air v
elo
city,
calculated fl
ow
Air v
elo
city,
calculated fl
ow,
air tempe
rature
Diameter
100 mm
60 mm
16 mm
Type of Measurement
Velocity
Vane
Vane
Vane
Temperature
Tc K
----
----
Tc K
----
Tc K
Measuring Range
Velocity
(m/s)
0.6…30
0.3…20
0.8…20
10…50
Temperature
-25…+80 -25…+80
(*)
-
25...+80
(*)
-
30..+120
(**)
-
25...+80
(*)
-
25…+80
Resolution
Velocity
0.01 m/s - 0.1 km/h - 1 ft/min - 0.1 mph - 0.1 knots
Temperature
0.1°C
----
---- 0.1°C ---- 0.1°C
Accuracy
Velocity
±(0.1 m/s
+1.5%f.s.)
±(0.1m/s
+1.5%f.s.)
±(0.2 m/s +1.0%f.s.)
Temperature
±0.1°C
----
---- ±0.1°C ---- ±0.1°C
Min. Velocity
0.6m/s
0.3m/s 0.8m/s 10m/s
Unit of Measurement
Velocity
m/s – km/h – ft/min – mph - knots
Flow
l/s – m
3
/s – m
3
/min – ft
3
/s – ft
3
/min
Duct Section for Flow Calcula-
tion
100…100000 cm
2
0.01…10 m
2
Calculation of Duct Section
(for Flow Measurement)
The area of the section can be assigned:
•
directly (cm
2
or inch
2
)
•
by setting the radium (cm or inch) for circular sections
•
by setting the side length (cm or inch) for rectangular sections
DO9847 Firmware Upgrade
Version 2.0 and following ones
Cable Length
~2m
(*) The indicated value refers to the vane operating range.
(**) Temperature limit refers to the probe where the vane and temperature sensor are located and not to the handle, the
cable and telescopic shaft which maximum working temperature is 80°C.