background image

17

Trigno

TM

 EMG System

Specifications

Typical Operating Range

(1)

20 m

RF Frequency Band

2400-2483 MHz (ISM band)

Power Consumption

<65 mW

Effective Radiated Power

9 mW

RF Protocol

Proprietary

Enclosure Dimension (main sensor)

27 x 37 x 15 mm

MiniHead Dimension

25 x 12 x 7 mm

Head Cable Length

200 mm

Full-charge Operation Time

(2)

7 hours (typical)

Recharge Time

(3)

<2.5 hours

Auto Shut-down timer

300 seconds

Temperature Range

(4)

5 - 40 degrees Celsius

Signal Range

11 mV (r.t.i)

EMG Channel Bandwidth

20 ± 5 Hz, >40 dB/dec

850 ± 100 Hz, >80 dB/dec

ACC Channels Bandwidths

DC - 50 ± 5 Hz, 20 dB/dec

EMG Channel Sampling Rate

2000 samples/sec, 1925.93 samples/sec*

Accelerometer  Sampling Rate

148.1 samples/sec/axis

EMG Channel Resolution Depth

16 bits

Accelerometer Resolution Depth

10 bits

EMG Channel Noise

(5)

<0.75 uV (rms), r.t.i.

Accelerometer Range

± 1.5g, ± 6g

Accelerometer Noise

0.007g (rms, ± 1.5g)

0.024g (rms, ± 6g)

Accelerometer Offset Error

±0.201g (XY, 1.5g)

+0.201g / -0.343g (Z, 1.5g)

±0.052g (XY, 6g)

+0.052g / -0.088g (Z, 6g)

(1)   Range is characterized in open office environments. Interfering RF sources in the 2.4  GHz spectrum, 

as well as absorptive objects occluding the RF communication path may degrade transmission dis-

tance. Stated range can be exceeded under favorable RF conditions.

(2) 

Battery duration is a function of charge and discharge conditions.  Optimal battery performance is 

obtained when the device is operated at room temperature.  Note that the stated Operation Time 

reflects the expected performance of a fully charged new battery used in a sensor that is transmitting 

data.  Operation Time is expected to decrease as a function of charge cycles, and when the sensor is 

searching for a network.

(3) 

80% of original battery capacity is maintained after 300 discharge/recharge cycles or after 2 years if 

recharge cycles are less than 300. These values represent typical expectations under normal condi-

tions. Actual performance will vary depending on usage conditions. 

(4) 

Operation beyond these temperature limits may damage the rechargeable battery.

(5) 

Input-referred noise is calculated as a root mean square over a 3 second window sampled at 1926 kHz.

Содержание TRIGNO

Страница 1: ...TRIGNOTM EMG System Copyright Delsys Incorporated Delsys Logo and EMGworks are Registered Trademarks of Delsys Inc MAN 015 1 3 Spring Contact Adapter User s Guide...

Страница 2: ......

Страница 3: ...Service and Support 4 Warnings and Precautions 5 Device Information 6 Disclaimer 8 System Requirements 8 Spring Contact Adpater Overview 9 Using the Sensors 10 Charging the Sensors 10 Sensor Pairing...

Страница 4: ...the EMG and supporting signals by a qualified individual is required Rx ONLY DO NOT USE on Patients with implanted electronic devices of any kind including cardiac pace makers or similar assistive de...

Страница 5: ...sitive electronic device Avoid static discharges Do not oper ate or store near strong electrostatic electromagnetic magnetic or radioactive fields Interference from external sources may decrease the s...

Страница 6: ...cell Storing or operating the device and conse quently the cell outside of this temperature range may compro mise the integrity and the safety features of the cell Device Information Complies with Re...

Страница 7: ...ntenna type and its gain should be so chosen that the equivalent isotropi cally radiated power EIRP is not more than that required for successful communication This equipment has been tested and found...

Страница 8: ...d upon theory of warranty contract tort or otherwise No representative agent or licensed practitioner is authorized to waive this disclaimer DELSYS INC makes no diagnosis or prescription by virtue of...

Страница 9: ...er contacts to establish a local electrical reference from which to measure the EMG signals detected by the spring contacts The sensor body includes an onboard 1 5g 6g accelerometer Spring Contact Adp...

Страница 10: ...dapter for your location Ensure that the Trigno sensors are properly fitted in the recharge pockets The sensor LEDs will illuminate to amber during charging and green when charge is complete The recha...

Страница 11: ...Gworks initiate pairing by right clicking the Trigno hardware icon in the system notification area and select ing the appropriate menu item Figure 3 Invoking the Pair command Right click on the Trigno...

Страница 12: ...is being paired for the first time with the base station the choice will read Use Default Calibration rather than Use Auto detected Calibra tion Figure 5 The calibration can be auto detected or manual...

Страница 13: ...all 4 contacts are visible and one 2 slot interface to the sensor head on the opposite side of the spring contacts 1 Identify the 2 slot adhesive interfaces to be used with the spring contact sensor...

Страница 14: ...the skin Ensure that the refer ence contact location is appropriately prepared for optimal sig nal quality Adhesive Sensor Interfaces are for single use only Immediately discontinue use if skin irrit...

Страница 15: ...and are quite soft Care should be taken to preserve the integrity of these con tacts Do not scrape or dent these contacts Handle the sensors with care do not drop them on the ground or step on them Th...

Страница 16: ...he battery Periodically charge the sensors at least once every 3 months to extend battery life Do not submerge the sensors in any liquid under any cir cumstance The sensors contain sensitive electroni...

Страница 17: ...Y 6g 0 052g 0 088g Z 6g 1 Range is characterized in open office environments Interfering RF sources in the 2 4 GHz spectrum as well as absorptive objects occluding the RF communication path may degrad...

Страница 18: ...Trigno Spring Contact Adapter User s Guide 18 This page has been intentionally left blank...