Fixed air inlets should be used to obtain best
result.
The calibration must be carried out with real
milk (during milking).
Note!
Note!
Note!
Note! Collect reference values for the Periodic
testing using test probe.
Installation of the bucket for calibration:
It is important to secure the CV is installed be-
tween the bucket and the milk line to avoid re-
maining vacuum on the milk bucket.
Remove the milk tube from the MM27BC and
attach on the bucket without stressing the
conductivity cable.
Add a new milk tube between the milk meter
output and the milk bucket.
No 5 Change diaphragm of the RS
No 5 Change diaphragm of the RS
No 5 Change diaphragm of the RS
No 5 Change diaphragm of the RS----valve
valve
valve
valve
Replacement parts or kits:
Art No: 85850901
Art No: 85531630
The article numbers above are kits for 1 and 10
pcs.
No 6 Replace rubber tubes
No 6 Replace rubber tubes
No 6 Replace rubber tubes
No 6 Replace rubber tubes
20
98
42
20
98
43
Frequency
4000h of operation
Max. interval
24 months
Estimated time:
5 min
Service type:
Replace
1.
1.
1.
1. Open by turning the cover by turning slightly
counterclockwise.
2.
2.
2.
2. Clean gently. Do not disconnect the cable.
3.
3.
3.
3. Replace the diaphragm.
Frequency
2000h of operation
DeLaval milk meter MM27BC
DeLaval milk meter MM27BC
DeLaval milk meter MM27BC
DeLaval milk meter MM27BC
161 Preventive maintenance
9 (22)
© DeLaval 2012.All rights reserved.