nents should be carried out to verify that 
there are no big changes of equipment 
installation or leakages.

The five values below represent milking 
point parameters that if significantly changed 
may affect the MM27BC bias and therefore 
the results of the milk yield measurements.
If the routine test should fail, the MM27BC 
meters must be recalibrated with milk.

Vacuum level:
If the vacuum level has changed by more 
than 2.5 Kpa since the installation test, 
the routine test fails.
Claw type:
If another significantly different claw is in-
stalled, the routine test will fail.
Milk tubes:

If the diameter of the milk tubes be-
tween milking cluster and the milk 
meter has been changed, the routine 
test fails.

If the length of the milk tube between 
milking cluster and the milk meter has 
been changed by more than +/- 10%, 
the routine test fails.

If the milk tube between MM27BC 
meter and milk pipe line has been 
changed in diameter or in length, the 
routine test fails.

The height of the MM27BC meter relative 
to the cow platform:
If this distance has been changed by 
more then 7 cm, the routine test fails.
Air inlet:
Increased air inlets or leakages will 
cause deviations in the reading of the 
MM27BC milk meter. If the air inlet has 
significantly changed by more than 10% 
compared with the installation test, for 
example by air leakages along the cluster 
to the MM27BC milk meter, the routine 
test fails. The air inlet can be observed 
by visual inspection and by measuring 
the air leakage using the VPR100.

DeLaval milk meter MM27BC

DeLaval milk meter MM27BC

DeLaval milk meter MM27BC

DeLaval milk meter MM27BC

161 Preventive maintenance

5 (22)

© DeLaval 2012.All rights reserved. 
Printout from Memoviewer 11/6/2012.

Содержание MM27BC

Страница 1: ...parts Replacement parts or kits or kits or kits or kits 1 Replace MM27BC gasket 2000h 12 months 91470502 2 Replace fat sampler parts 2000h 12 months 85132601 3 Periodic checking and hints for sampling...

Страница 2: ...asket tightly into the cavity of the inlet tube with the correct orientation to match retangular channel It is important to press the inlet tube tight against the milk meter to compress the gasket as...

Страница 3: ...ings and a rubber gasket or sampler plug that must be replaced once a year 2 2 2 2 The silicon tubes and rubber bottle holder must be replaced before they have deteri orated typically 1 2 years Freque...

Страница 4: ...ion MM27BC visual inspection MM27BC visual inspection MM27BC visual inspection Milking parlour parameters Milking parlour parameters Milking parlour parameters Milking parlour parameters Fig 1 Test pr...

Страница 5: ...g cluster and the milk meter has been changed by more than 10 the routine test fails If the milk tube between MM27BC meter and milk pipe line has been changed in diameter or in length the routine test...

Страница 6: ...a meters Check of internal parameters The meters Optical channel must be checked The milk channel characteristics and optical field must stay the same as it was at the time for calibration The test pr...

Страница 7: ...t fails the operator can try to fix the problem by using the Psion RECAL option If still the MM27BC self test fails the MM27BC meter is defective and should be re placed 2 0 9 8 4 0 2 0 9 8 4 1 v m x...

Страница 8: ...lower number is the minimal flow and the higher the maximal flow measured The numbers shown are 4 digits and usually be tween 3 000 and 4 000 kg per minute 3 9 3 9 3 9 3 9 Write the minimal and maxima...

Страница 9: ...the RS No 5 Change diaphragm of the RS No 5 Change diaphragm of the RS valve valve valve valve Replacement parts or kits Art No 85850901 Art No 85531630 The article numbers above are kits for 1 and 1...

Страница 10: ...imated time 10 min Service type Replace 1 1 1 1 2 2 2 2 Replace tubes with the service items Frequency 4000h of operation Max interval 24 months Estimated time 10 min Service type Replace DeLaval milk...

Страница 11: ...e 91677780 2 2 2 2 2 2 2 2 Test Procedure Test Procedure Test Procedure Test Procedure 1 1 1 1 Find kits that include silicon tube service items in original parts 2 2 2 2 Replace tubes with the servic...

Страница 12: ...o menu for MM25W MM27BC Choose ALCOM Use the VPR100 to read all Alcom node number in the parlour Mark node number on a piece of Tape on each milkmeter Make sure no milkmeter has the same number If the...

Страница 13: ...d in the right order This is done by going into calibration menu press hash key followed by pressing on the Alcom number that ap pears default 80 Enter the MM27BC Alcom number and press return Press o...

Страница 14: ...Laval milk meter MM27BC DeLaval milk meter MM27BC DeLaval milk meter MM27BC DeLaval milk meter MM27BC 161 Preventive maintenance 14 22 DeLaval 2012 All rights reserved Printout from Memoviewer 11 6 20...

Страница 15: ...Laval milk meter MM27BC DeLaval milk meter MM27BC DeLaval milk meter MM27BC DeLaval milk meter MM27BC 161 Preventive maintenance 15 22 DeLaval 2012 All rights reserved Printout from Memoviewer 11 6 20...

Страница 16: ...Laval milk meter MM27BC DeLaval milk meter MM27BC DeLaval milk meter MM27BC DeLaval milk meter MM27BC 161 Preventive maintenance 16 22 DeLaval 2012 All rights reserved Printout from Memoviewer 11 6 20...

Страница 17: ...Laval milk meter MM27BC DeLaval milk meter MM27BC DeLaval milk meter MM27BC DeLaval milk meter MM27BC 161 Preventive maintenance 17 22 DeLaval 2012 All rights reserved Printout from Memoviewer 11 6 20...

Страница 18: ...Laval milk meter MM27BC DeLaval milk meter MM27BC DeLaval milk meter MM27BC DeLaval milk meter MM27BC 161 Preventive maintenance 18 22 DeLaval 2012 All rights reserved Printout from Memoviewer 11 6 20...

Страница 19: ...reference values shall be recorded at the time for the installation test just after the DeLaval milk meter MM27BC DeLaval milk meter MM27BC DeLaval milk meter MM27BC DeLaval milk meter MM27BC 161 Prev...

Страница 20: ...by pressing tools on the VPR 100 menu or on the PSION press 8 TECHM and then 1 PARMS Then insert the test probe in the out let of the meter After a few seconds the meter should display PASS on the dis...

Страница 21: ...be length 10 MM27BC height 10 cm Air inlet 10 Mp Mp Mp Mp Ref Ref Ref Ref bias bias bias bias Curr Curr Curr Curr bias bias bias bias Ref Ref Ref Ref sub soft sub soft sub soft sub soft Ref Ref Ref Re...

Страница 22: ...Laval milk meter MM27BC DeLaval milk meter MM27BC DeLaval milk meter MM27BC DeLaval milk meter MM27BC 161 Preventive maintenance 22 22 DeLaval 2012 All rights reserved Printout from Memoviewer 11 6 20...

Отзывы: