4
INPUT HEADER ADJUSTMENT
Defi ne Instruments recommends that you specify your sensor type(s) when you
place your order, to avoid unnecessary removal of the input module.
4.1 - Input header settings
The analog input board for the PRO–CTR has adjustable headers for Sensor Type (A),
Noise Filtering (B) and Excitation Voltage (C). Refer to the tables below to determine
whether the
default header positions
(shown in black) are suitable for your applica-
tion. If required, follow the instructions in 4.2 to adjust the header positions.
Quadr
atur
e IP
234K 2006
©
Off
On
NC
NPN
PNP
TT
L
24V 5V
CH1
CH3
CH2
COUNT
FREQ
A
B
C
D
Sensor Type (A)
CH1= Primary Ctr, CH2= B Ctr, CH3= C Ctr
NC
Not Connected
NPN
NPN Sensor (Sink)
PNP
PNP Sensor (Source)
TTL
TTL Sensor
Excitation Voltage (C)
24V
Standard setting
5V
May apply to some encoders
Noise Filtering (B)
CH1= Primary Ctr, CH2= B Ctr, CH3= C Ctr
Off
20kHz, for high-speed counting
On
1kHz low pass fi lter, for a noisy
signal or mechanical contact
Mode Header (D) -
Do not adjust!
COUNT
Always use this setting
FREQ
Not used for PRO–CTR