message string. If '
$
' is used, a 50ms minimum delay is inserted before a reply is
sent. If '
*
' is used, a 2ms minimum delay is inserted before a reply is sent.
Custom ASCII Read/Write Examples
Example
Description
SR$
Read display value from all controllers, 50ms delay.
S15R$
Read display value from controller address 15, 50ms delay.
S3U40*
Read unformatted data in channel 4 from controller address 3, 2ms delay.
S2W2 -10000$
Write -10000 to the display register of controller address 2, 50ms delay.
SWT CHAN_1$
Write ASCII text string Chan_1 to channel 1 text register, 50ms delay.
8 Bit Unsigned
8207
Baud rate
8211
Serial address
8215
Serial mode
16 Bit Unsigned
4181
Hysteresis SP 1 (Batch SP)
4182–4185
Hysteresis SP 2
–
5
4197
Make delay SP 1 (Batch SP)
4198–4201
Make delay SP 2
–
5
4213
Break delay SP 1 (Batch SP)
4214-4217
Break delay SP 2
–
5
5173
Batch count increment
24 Bit Signed (2 x 16 Bit)
2509
Load value (Primary)
2511
Load value (B counter)
2513
Load value (C counter)
32 Bit Signed (2 x 16 Bit)
7
Batch result
9
Rate
13
Primary counter
15
B counter
11
C counter
37
Batch tare
39
Batch count
57
Peak
59
Valley
111
Setpoint 1 (Batch setpoint)
113
Setpoint 2
115
Setpoint 3
117
Setpoint 4
119
Setpoint 5
239
Alarm status
Custom ASCII Registers -
Active for models with relay output installed
Controller Response -
Aft er the controller has completed a read or write instruction,