Primary Counter
Sig A
Sig B
Primary
counter
increments
:
On a
rising
edge A signal when B is
low
Primary Counter
Sig A
Sig B
Primary
counter
decrements
:
On a
rising
edge A signal when B is
high
6.8 - Count
This mode is particularly useful for multi-input systems.
Signal C
Primary
Counter
Primary
counter
increments
:
On a
rising
edge C signal only
6.9 - RSOFAB - Reset to off set A, B mode
In this mode the
primary
and
secondary
counters are reset to the values stored in
the meter's reset off set registers, and they continue counting aft er being reset. This
is useful for setting a position to a known reference position.
Signal C
Index/Marker
Primary
reset offset
register
Secondary
reset offset
register
Primary Counter
Continues counting
afer being reset
Secondary Counter
Continues counting
afer being reset
RESET
RESET
On a
falling
edge C signal:
Primary
counter is reset to the value stored in
primary reset offset register
Secondary
counter is reset to the value stored
in
secondary reset offset register
Both counters
continue to count after being
reset