12
If any probe other than the
PosiTector DPM
probe is
attached to the
PosiTector Advanced
body while connected to
a
PosiTector CMM IS
probe, the gage body will ignore the
attached probe and only display the
PosiTector CMM IS
measurements.
Recorded measurement values can be reported using PosiSoft
Solutions. See the
PosiTector DPM
Instruction Manual for more
information.
Step 6 – Remove the PosiTector CMM IS Probes
The
PosiTector CMM IS
probes are designed to be reusable.
Once measurements are complete, use the extraction tool to
remove the probes from the holes. Clean any visible debris or
concrete dust from the probes and inspect the silicone fins. If the
silicone fins are damaged or worn, they should be replaced. If no
further tests are required, remove the batteries by unscrewing the
battery covers and pushing the battery out using the probe
extraction tool and place each probe into a Calibration Check
Chambers for storage.
Once testing is complete, the holes should be filled with a
cementitious patching compound in accordance with ASTM
F2170.
NOTE:
Specifications
Range
Accuracy
Resolution
Temperature
0˚ to 80˚ C
± 0.5˚ C
0.1˚ C
32˚ to 175˚ F
± 1˚ F
0.1˚ F
Humidity
10 to 90%
± 2%*
0.1%
> 90%
± 3%*
* 0 – 65˚ C (32 – 150˚ F)
Содержание PosiTector CMM IS
Страница 2: ......
Страница 17: ...Manual de instrucciones v 1 0 Medidor de humedad de concreto CMM IS...
Страница 32: ...Gebrauchsanweisung v 1 0 Station res Betonfeuchtigkeits Messger t CMM IS...
Страница 47: ...Notice d instruction v 1 0 Instrument de mesure d humidit du b ton CMM IS...
Страница 62: ......
Страница 63: ......