Step 3. Unplug the sensor module
Step 4. Plug the new sensor module into the PCB
Step 5. Insert the batteries and start calibration of the
.
Step 6. Place the filter back.
This GHC gas sensor module has the default configuration, however, those parameters can be changed. The user can
change the configuration on the wireless transmitter so that the complete sensor (tran wireless) delivers the
proper output value. Below are some configuration parameters that store in the flash memory of the wireless
transmitter.
Description
Unit
Default
Format Property
Comment
CONSTANT_A
1
Float
R/W
Constant a for scaling measured value. This value
would be changed after calibration.
CONSTANT_B
0
Float
R/W
Constant b for scaling measured value. This value
would be changed after calibration.
HIGH_CUT
1E+09
Float
R/W
High cut value for scaled_value
LOW_CUT
0
Float
R/W
Low cut value for scaled_value
SENSOR RESPONSE
TIME
S
100
uint16
R/W
* Do not change this value
C_H_FACTOR
4.4
Float
R/W
4.4 for CH4, 1.7 for C3H8
END.
8. Default configuration
Содержание Lidar People Counter
Страница 2: ...Daviteq Optical Sensing Technologies...
Страница 8: ...Daviteq Radar Sensing Technologies...
Страница 9: ...Daviteq Gas Detection Technologies...
Страница 27: ......
Страница 34: ......
Страница 38: ...C m u n i M12 v o c m bi n Si t ch t k t n i M12...
Страница 43: ...Daviteq Electrical Sensing Technologies...
Страница 51: ...Daviteq Ultrasonic Level Distance Measuring Technologies...
Страница 57: ...Thermal mass flow Pitot tube Orifice plate Daviteq Air Gas Flow Measurement Technologies...
Страница 64: ...Figure 12 Adjust the orientation of the probe 5 7 Attach the Wireless Transmitter to the probe...
Страница 67: ...Daviteq Inertial Measurement Technologies...
Страница 80: ...Daviteq Pressure Measurement Technologies...
Страница 87: ...Tel 84 28 6268 2523 4 ext 122 Email info daviteq com www daviteq com...
Страница 88: ...Daviteq Soil Measurement Sensors...