FUNCTIONAL DESCRIPTION
2.1
Measurement principle
NIBP (Non-Invasive Blood Pressure) is an indirect method for measuring blood pressure.
The NIBP measurement is performed according to the oscillometric measuring principle. The cuff is
inflated with a pressure slightly higher than the presumed systolic pressure, and deflated at a
speed based on the patient’s pulse, collecting data from the oscillations caused by the pulsating
artery. Based on these oscillations, values for systolic, mean, and diastolic pressures are
calculated.
The following parts are necessary for NIBP measurement:
•
NIBP module
•
Twin hose (adult or infant model)
•
Blood pressure cuffs (different sizes)
2.2
Main components
The NIBP module consists the following parts:
•
NIBP board
•
Pneumatics and hosing
•
NIBP air pump
•
Zero valve
•
Check valve
•
Bleed valve
•
Exhaust valves (2)
•
Pressure transducers (2)
•
Module keyboard and status indicator LEDs
•
Front panel keys: Auto On/Off, Set Cycle Time, Stat On/Off, Start/Cancel
Auto
On/Off
Set
Cycle
Time
STAT
On/Off
NIBP
Auto
Manual
STAT
Measuring
Start
Cancel
Figure 2
Front panel of NIBP Module
Содержание S/5 M-NIBP
Страница 2: ......
Страница 36: ...Datex Ohmeda S 5 monitors Document No 800 1015 1 32 This page intentionally left blank ...
Страница 44: ...Datex Ohmeda S 5 monitors Document No 800 1015 1 40 ...
Страница 45: ...APPENDIX A NIBP Module M NIBP Document No 8001015 1 41 APPENDIX A ...
Страница 46: ...Datex Ohmeda S 5 monitors Document No 800 1015 1 42 ...