![Datex-Ohmeda S/5 M-NIBP Скачать руководство пользователя страница 47](http://html.mh-extra.com/html/datex-ohmeda/s-5-m-nibp/s-5-m-nibp_technical-reference-manual-slot_513844047.webp)
APPENDIX A, Service check form, NIBP Module, M-NIBP
A-1(2)
Document No. 8001015-1
SERVICE CHECK FORM
Non-Invasive Blood Pressure Module, M-NIBP
Customer
Service
Module type
S/N
Service engineer
Date
OK = Test OK
N.A. = Test not applicable
Fail = Test Failed
OK
N.A.
Fail
OK
N.A.
Fail
1.
Internal parts
2.
External parts
3.
Pump’s filter
4.
Installation
Notes
5.
Recognition
6.
Module software
NIBP
7.
Communication and
memories
8.
LEDs and membrane
keys
9.
Pump and valves
10.
Leak test
≤
5 mmHg/min
11.
Calibration check
Measured B1
Measured B2
Allowed range
0 mmHg
±
10 mmHg
100 mmHg
100 + z.o.
±
2 mmHg
200 mmHg
200 + z.o.
±
3 mmHg
260 mmHg
260 + z.o.
±
4 mmHg
z.o. = zero offset at 0 mmHg pressure
12.
Watchdog timer activation pressure
9...19 / 5...10 mmHg
Содержание S/5 M-NIBP
Страница 2: ......
Страница 36: ...Datex Ohmeda S 5 monitors Document No 800 1015 1 32 This page intentionally left blank ...
Страница 44: ...Datex Ohmeda S 5 monitors Document No 800 1015 1 40 ...
Страница 45: ...APPENDIX A NIBP Module M NIBP Document No 8001015 1 41 APPENDIX A ...
Страница 46: ...Datex Ohmeda S 5 monitors Document No 800 1015 1 42 ...