NIBP Module, M-NIBP
Document No. 800 1015-1
31
Watchdog BEEP
Connect manometer to the front panel and pump pressure into the module. When the AD5 value
changes from negative to positive value (at about 5 mmHg) a beep is heard. This is the watchdog
threshold pressure. Beyond this pressure the watchdog is active and cut pressures at about 2 min.
(adult).
5.1.7
NIBP Watchdog menu
NOTE: Parameter values in Service Data are for reference only.
Test ADULT
Test ADULT
is to test watchdog timer in adult mode (120 to 140 seconds).
Test INFANT
Test INFANT
is to test watchdog timer in infant mode (about 60 to 70 seconds).
Stop Test
Stop Test is for stopping the test.
Watchdog Data
See NIBP Service menu in chapter 5.1 for information on general items
Pressure,
Zero, Protect
handle,
Calibr. prot., +15 V power, AD0
to
AD7
as well as
Timeouts etc.
Watchdog Interval
shows the time the +15 Vdirty stays on during the test.
Adult watchdog time testing
Select Test ADULT and press the ComWheel. Watchdog interval starts counting up seconds and
keeps on counting as long as the +15 Vdirty is on. The time should be 120 to 140 seconds.
Infant watchdog time testing
Select Test INFANT and press the ComWheel. Watchdog interval starts counting up seconds and
keeps on counting as long as the +15 Vdirty is on. The time should be 60 to 70 seconds.
Содержание S/5 M-NIBP
Страница 2: ......
Страница 36: ...Datex Ohmeda S 5 monitors Document No 800 1015 1 32 This page intentionally left blank ...
Страница 44: ...Datex Ohmeda S 5 monitors Document No 800 1015 1 40 ...
Страница 45: ...APPENDIX A NIBP Module M NIBP Document No 8001015 1 41 APPENDIX A ...
Страница 46: ...Datex Ohmeda S 5 monitors Document No 800 1015 1 42 ...