Detecting arrhythmia alarms
NOTE: A clinician must analyze the arrhythmia information in conjunction with other clinical findings.
For details about detection performance and test results of the arrhythmia algorithm testing, please refer to “User’s Reference Manual: ECG.”
Alarm Criteria
Asystole
Cardiac arrest, no QRS complexes for five seconds
Bradycardia
HR below the HR alarm limit
Frequent PVCs
PVCs per minute above the alarm limit
Missing Beat
Actual RR rate interval more than 1.8 times the average RR interval
Multif. PVCs
Over the last 15 beats two or more premature ventricular beats with different morphologies are detected
R on T PVC
Early PVC, beat detected as a PVC and previous beat as not a PVC; current RR interval is less than half of the previous
RR interval
Tachycardia
HR over the HR alarm limit
Ventricular bigeminy
The following pattern is detected: N, V, N, V, N, V where N = normal, V = PVC (every other beat is a PVC)
Ventricular couplet
Two consecutive ventricular beats, preceded and followed by a normal beat
Ventricular fibrillation
Fibrillatory waveform caused by ventricular fibrillation
Ventricular run
Three or more consecutive PVCs and rate of successive beats over 120 bpm
Ventricular tachycardia
Six or more consecutive PVCs and rate of successive beats over 120 bpm
Ventricular trigeminy
The following pattern is detected: N, N, V, N, N, V, N, N, V where N = normal, V = PVC (i.e., every third beat is a PVC)