50
50
Before cleaning
1. Turn off the monitor from the ON/standby key.
2. Disconnect the power cord.
After cleaning
1. Let dry completely.
2. Reconnect the power cord and turn on the monitor.
More comprehensive checking
See the “Technical Reference Manual.”
WARNING: After cleaning or if liquid has accidentally
entered the monitor, ensure that every part of the
monitor is dry before reconnecting it to the power
supply.
Water trap in the N-FCREC and N-FC modules
•
Empty the container whenever half full.
•
Change the water trap every two months and when the text
‘Replace D-Fend’ appears.
•
The water trap cartridge is disposable. Do not wash or reuse the
cartridge.
Other accessories
For information on how to clean and check reusable accessories, see
the accessory package or the “Supplies and Accessories” catalog. Do
not reuse single-use disposable accessories.
Conditioning the batteries
Condition batteries regularly to maintain their useful life. Condition a
battery every six months or when the message ‘Replace Battery A’
appears on the screen. Always observe the messages and symbols on
the screen to see the battery status. You can also check the status
through
Monitor Setup
–
Battery Setup
. For more information, see
sections "Replacing the batteries", "Symbols" and "Messages."
Conditioning a battery is best done on an external charger. To do this,
please, refer to the "User’s Reference Manual."
Power interruption
NOTE: Always use the S/5 FM with batteries inserted. Otherwise all
trend data, snapshots and temporary settings are lost if the power
cable is detached from the mains.
If the monitor is in standby or turned off, trend data and the latest
user-made settings remain in the monitor memory for 15 minutes
even if the mains power is interrupted. After 15 minutes, trend data is
lost and the monitor returns to the user default settings (start-up
mode).
Changing fuses
1. Remove the power cord if used.
2. Remove the fuse holder by pushing the locking pin and pulling the
holder gently out, see also page 10.
3. If a fuse is blown, replace it with a fuse of the correct type and
rating.
Содержание S/5 FM
Страница 1: ...S 5 FM User s Guide guide_FM_uusi pmd 03 04 2004 8 31 1 ...
Страница 22: ...20 20 For your notes ...
Страница 32: ...30 30 For your notes ...
Страница 60: ...58 58 For your notes ...