28
28
During monitoring
To remove a patient temporarily from the monitor, use standby:
1. Disconnect patient cables or the measurement module and
ensure that the monitor receives no vital signs and that the NIBP
autocycling is turned off.
2. Press the
Admit/Discharge
key.
3. Select
Standby
and one of the options.
4. Accept with
YES
. Alarms are silenced and all the gathered data is
saved. New data is not collected.
Retrieving information after standby
If the monitor has been in the standby state:
1. Reconnect patient cables or the measurement module.
2. Press any key or push the ComWheel.
3. Select
YES
in the opened
Contin. Previous
window.
Alarms are activated. The previous data is retrieved and the monitor
starts collecting new data. NOTE: Zero invasive pressure lines.
Ending monitoring
1. Print necessary data: press the
Print/Record
key.
2. Wait until the printing is finished. Then clear the patient data and
return the settings, including alarm limits, to their defaults through
Admit/Discharge
-
Discharge - YES.
3. Turn off the monitor from the ON/standby key if the monitor will
not be used.
4. Clean the monitor according to the instructions.
Automatic discharge of the patient
The monitor discharges a patient automatically after 24 hours when
vital signs for some parameters (ECG, Art, NIBP, SpO
2
, Resp and CO
2
(with N-FCREC and N-FC)) are not available. When this happens, all
trend data will be cleared and alarm limits set to default values.