Unable to measure Sys
NIBP: Systolic pressure probably higher than maximum inflation pressure or artifacts interfere in the
systolic area.
Unstable zero pressure
NIBP: Pressure is unstable at start of the measurement. Calm the patient and retry.
Weak pulsation
NIBP measuring problem:
•
Improper cuff position or attachment.
•
Weak or abnormal blood circulation.
•
Slow heart rate associated with artifacts.
•
Patient is moving
•
Air leakage.
… high/… low
Measured value (…) exceeds the alarm limit. Check patient condition. Adjust alarm limits.
… Lead off
One of the ECG lead wires (…) is off.