ALARMS
After the monitor is turned on or after the case is reset, the alarm limits are activated only when the physiological signals have been
inside the alarm limits for 15 seconds. To enable the alarms, connect the patient cables. Alarms are operative even when the
measurement is not selected on the screen (except for the respiration measurement), unless the source is selected
OFF
.
1
2
3
mm
1mV
1mV
II
V5
aVL
0.9
0.0
0.0
58
0.9
When an alarm becomes active:
(1) Messages appear in order of priority.
(2) The measurement value and the alarm
LED flash. The background color
signifies the alarm category.
(3) A message gives more detailed
information in some cases.
An audible alarm sounds.
Tachy
ECG
H
R
Pacer
hidden
/min
Learning
S
T
Alarm categories
The priority depends primarily on the cause and alarm duration.
Visual
Meaning
Tone pattern (selected when the system is configured)
Red
For life threatening situations
double beep every 5 seconds or continuous beeping:
– – – – – 5 – – – – – / – – – – –
Yellow
For serious but not life threatening
problems
Triple beep every 19 seconds or double beep every 5 seconds:
– – – 19 – – – / – – 5 – –
White
Advisory
Single beep: –
– 13 –