Message
Explanation
Check stim. electrodes
NMT stimulus current could not be delivered due to poor stimulus electrode connection or damaged
cable.
Cuff loose
Cuff occlusion
Cuff overpressure
NIBP:
•
Cuff is not attached to patient.
•
Cuff is too loose or hose is not connected.
•
Tubes or hose are kinked.
•
Cuff is squeezed during measurement.
Check NIBP cuff hose and tubes and restart measurement.
EEPROM error
Memory checking failed. Contact authorized service personnel.
EMG electrodes off
NMT: EMG recording electrodes are off.
Infl. limits! Check setup
NIBP: Adult or child cuff is used but the infant mode has been selected.
Leads off
ECG trunk cable, all lead wires, or neutral electrode (RL/N) are disconnected. ECG inoperative (offset
voltage between two electrodes is too high). Possible during defibrillation.
Measurement off
NMT cable is connected but measurement has not started.
MVexp << MVinsp
Exhaled volume is markedly smaller than inhaled. Check the whole system for leakage.
MVexp < 0.5l/min
(with Pedi-lite:
MVexp < 0.2l/min)
Measured volumes are too small for reliable calculation and, for example, waveforms and loops may be
unreliable.
NIBP manual
Autocycling mode is interrupted because of air leakage or loose cuff. Check setup and restart autocycling.
No P1 transducer
Invasive blood pressure transducer or cable of channel P1/P2 not connected.
No SpO2 probe
Check SpO2 sensor connection.
No SpO2 pulse
SpO2 pulse signal is poor. Try other measuring sites.
Noise
ECG: Unreliable HR calculation or distorted waveform; possible during diathermy.
Performing temp test
Monitor makes a two point calibration test for temperature immediately after the warm-up time
and after that once in 10 minutes. Test lasts 10 seconds.