WARNING: The impedance respiration measurement may cause rate
changes in Minute Ventilation Rate Responsive Pacemakers. Set
the pacemaker rate responsive mode off or turn off the impedance
respiration measurement on the monitor.
NOTE: A clinician must analyze the ST segment changes in
conjunction with other clinical findings.
Monitoring Pacemaker Patients
1. Press
the
ECG
key.
WARNING: Do not rely entirely upon rate meter alarms when
monitoring patients with pacemakers. The monitor may count the
pacemaker pulses as heartbeats. In this case, asystole and
ventricular fibrillation may go undetected. Always keep these
patients under close surveillance and monitor their vital signs
carefully.
2. Select
ECG Setup - Pacemaker
and select one of the
following:
•
Show
= Pacemaker spike is displayed on ECG.
•
Sensit
= Sensitive pacemaker detection; spike displayed
on ECG
•
ON R
= Pacemaker suppression weakened; asystole
alarm may not be reliable with active pacemakers.
•
Hide
= Pacemaker spike is not displayed on ECG.
NOTE: Pacemaker detector may not operate correctly during the
use of high-frequency (HF) surgical equipment. The
disturbances of HF surgical equipment typically cause false
positive pacer detection.