DT1310 xx xx
26
20220128-V0
7.8. Analogue output scaling
Function
There are two output versions of the instrument:
DT1310 xx Ix xx: 0 / 4-20 mA current output or
DT1310 xx U xx: 0 / 2-10 V voltage output
It can be assign any temperature or resistance range or
potentiometer scale to the output.
[Factory default: 4 mA = -200.0 °C and 20 mA = +800.0
°C.]
1. Type-in the value of the start point of the range or scale
(0 / 4 mA or 0 / 2 V).
2. Type in the value of the end point of the range or scale
(20 mA or 10 V).
The figure shows the factory default settings.
Sequence of operations
Note:
It can be used either dot (.) or comma (,) for decimal.
When you type in more decimals the software corrects it for
one decimal after clicking on the “download settings”
button.