23
7.6. Setting averaging
Function
The instrument performs cca. 12 measurements in each
seconds. (In 3 wire mode only 5 measurements). The
measurement result is generated as the average of several
measurements.
Here you can define the number of measurements that
should be used for calculating the averaged numerical
value. By increasing this number the signal stability
increases, but the signal settling time becomes lower.
[Default factory setting: 16]
1. Select the appropriate output signal by clicking the
button.
The figure shows the factory default setting.
Sequence of operations
Note: Selecting the „1” value the last measured result will
be transmitted. Measured results - measured before the
last measurement - have no effect on the output signal.
Settling time after stabilizing the input signal (4 wire mode):
Averaging number:
Settling time (sec):
1 0.1
2 0.2
4 0.4
8 0.7
16 1.4
32 2.7
64 5.4
Содержание DT1310 Series
Страница 1: ...DT1310 xx xx Intrinsically Safe Temperature Resistance Transmitter Operating Instructions...
Страница 41: ...DT1310 xx xx 20220128 V0 41 10 2 Application example...
Страница 42: ...DT1310 xx xx 42 20220128 V0 10 3 ATEX Tan s tv ny...
Страница 43: ...DT1310 xx xx 20220128 V0 43...
Страница 44: ...DT1310 xx xx 44 20220128 V0...
Страница 45: ...DT1310 xx xx 20220128 V0 45...
Страница 46: ...DT1310 xx xx 46 20220128 V0...
Страница 47: ...DT1310 xx xx 20220128 V0 47...
Страница 48: ...DT1310 xx xx 48 20220128 V0...
Страница 49: ...DT1310 xx xx 20220128 V0 49...
Страница 50: ...DT1310 xx xx 50 20220128 V0...
Страница 51: ...DT1310 xx xx 20220128 V0 51...
Страница 52: ......