DT1010 xx xx
20
20180403-V1
7.4. Analogue output setting
Function
The output signal can be current: 0 / 4-20 mA, or
voltage 0 / 2-10 V. (should be specified at order)
[Factory default: 4-20 mA]
1. Select the appropriate output signal by clicking the
button.
The figure shows the factory default setting.
Sequence of operations
Note:
It is recommended to use 4-20 mA.
It’s advantages:
•
wire break can be detected by the signal processing unit.
•
it can drive a loop-indicator (e.g. DT4200)
7.5. Error current / voltage setting
Function
In a case of an error state the output current / voltage is
forced into error state. The value of the error current /
voltage can be set:
Current: 3.2 mA or 21 mA
Voltage: 1.6 V or 10.5 V
[Factory default: 3.2 mA]
1. Select the appropriate output signal by clicking the
button.
The figure shows the factory default setting.
Sequence of operations
Note:
In a case of 0-20 mA output current the error current is 21
mA, independent from setting.
Содержание DT1010 Series
Страница 1: ...DT1010 xx xx Temperature Resistance Potentiometer Transmitter Operating Instructions...
Страница 38: ...DT1010 xx xx 38 20180403 V1 10 2 Application example...
Страница 39: ...DT1010 xx xx 20180403 V1 39...
Страница 40: ......