76
M-Class
Advanced Entry Calibration (continued)
Step
Operator Action
Displayed Message
Comment
13
Using the
FWD
Key, select the
Gain Number determined in Step
12. Press the
Key to
select the setting.
!
HBJO!USBO!!!!!!=zzz?+1
9!!!!!!!!!)1!.!42*!
!
Or, for reflective media:
HBJO!SFGM!!!!!!=zzz?+1
9!!!!!!!!!)1!.!42*!
Selection is denoted with an ‘*’.
14
Place the media in the Media
Sensor. Record the sensor
reading and label it ‘P’
(paper).
Place the backing, mark, or
notch in the Media Sensor.
Record the sensor reading and
label it ‘G’ or ‘M’ (Gap or
Mark).
Remove all media. Record
the sensor reading and label it
‘E’ (Empty).
!
HBJO!USBO!!!!!!=zzz?+1
9!!!!!!!!!)1!.!42*!
!
Or, for reflective media:
HBJO!SFGM!!!!!!=zzz?+1
9!!!!!!!!!)1!.!42*!
Where ‘yyy’ is a numerical value representing the
current sensor reading.
15
Press the
Key. Then
press the
FWD
Key.
!
BEWBODFE!FOUSZ!
TFOTPS!MFWFMT!
!
The sensor readings must be manually entered into
the printer now.
Содержание M-Class
Страница 1: ...M Class Operator s Manual ...
Страница 24: ...16 M Class ...
Страница 46: ...38 M Class ...
Страница 86: ...78 M Class ...
Страница 96: ...88 M Class ...
Страница 108: ...100 M Class ...
Страница 122: ...114 M Class ...
Страница 128: ...120 M Class Network Status ...
Страница 129: ...M Class 121 TCP IP Configuration ...
Страница 131: ...M Class 123 System Settings Media Settings and Print Control continued Printer Options Communications and Diagnostics ...
Страница 132: ...124 M Class Printer Options Communications and Diagnostics continued ...
Страница 133: ...M Class 125 Network Print Options Reset Network Parameters ...
Страница 134: ...126 M Class Change Password ...
Страница 154: ...146 M Class ...
Страница 160: ...152 M Class ...
Страница 166: ...158 M Class ...