manualshive.com logo in svg
background image

User guide 

Ultrasonic heat meter SonoMeter 30

2 | VU.SH.I1.02

© Danfoss | 2019.10

1. General information

This instruction designed for ultrasonic heat 

meter SonoMeter 30 flow and energy verification 

and calibration.
Heat meter energy measurement error (δ

Es

assess the application of the partial verification 

method: assesses the calculator thermal energy 

measurement error (δ

E

), the flow volume 

measurement error (δ

V

) and temperature sensor 

temperature difference measurement error (δ

T

). 

Heat meter energy measurement error (δ

Es

) are 

all components of the arithmetic sum of the 

errors:

Es

 

=

 

׀

 

E

׀

 

 

+

 

׀

 

T

׀

 

1)  Energy pulse out terminals (Fig. 1) connected 

to the pulse counter (not required when 

reading the measurement results directly 

from the display or via a digital interface)

Pin 53

 - Energy Pulse output (Pulse)

Pin 52

 - Energy Pulse output (GND)

2. Test instruction

2.1 Calculator energy 

measurement error (δ

E

) test

2)  Temperature sensors placed in the 

temperature testing bench (for temperature 

difference simulation).

3)  Verification mode is activated by connecting 

TEST jumper contacts 1 (FIg. 1).

4)  Determination errors:

- before each measurement run automatic 

verification mode with long button clicking 

on the meter (simulated flow pulses),

- When the LCD display stops blinking “TEST”, 

energy and volume values can be read 

directly from the LCD screen or through the 

optical / digital interface. The pulses shall 

begin calculated with the corresponding 

nominal flow pulse values as follows:

Permanent flow q

p

(m

3

/h)

Energy pulse value,

(Wh/imp)

0.6

0.1

1.0

0.2

1.5

0.2

2.5

0.5

3.5

0.5

6.0

1

10.0

2

15.0

5

- energy E (kWh) measurement error is 

calculated by formula:

E

 = (E – E

0

) / E

0

 • 100 %

 

 

there:

 

E

0

 – the conventional true energy E value 

(kWh), calculated in consideration of the 

fluid temperature values simulated Θ1, Θ2 

and simulated change in volume values:

Flow sensor mounting 

place

E

0

 calculating formula

Supply pipe

E

0

=V1 • ρ1 • (h

T1

-h

T2

)

Return pipe

E

0

=V1 • ρ2 • (h

T1

-h

T2

)

There:

E

0

 – the estimated true value of the energy component 

(kWh);

V1

 – simulated water volume value (m

3

);

ρ1…ρ2

 – density of water, corresponding to the sumulated 

temperature value Θ1,Θ2, (kg/m

3

);

h

T1

… h

T2

 – relative enthalphy of water, corresponding to the 

simulated temperature value Θ1,Θ2

1)  Volume pulse output terminals (Fig. 1) 

connected to the pulse counter (not required 

when the measurement results are directly 

reading from the display or via a digital 

interface)

Pin 51

 - Volume Pulse output (Pulse)

Pin 50

 - Volume Pulse output (GND)

2.2 Volume measurement 

errors (δ

V

) test

2) Switched on verification mode by connecting 

TEST jumper contacts 1 (1 pav),

3) The meter mounted on the flow test bench, 

flow are simulating.

Recommended minimum pulse content and 

measurement time:

Flow

Pulse content

Measurement time

0.1 q

p

 <q ≤ q

s

N≥1000

T> 1 (min)

q

i

 ≤ q ≤ 0.1 q

p

N>500

T > 1 + 8 q

i

/q, (min)

Fig. 1

Отзывы: