Ultrasonic heat meter SonoMeter 30

VU.SH.I1.02 | 3

© Danfoss | 2019.10

It must be maintained both minimum time and 

minimum pulse quantities. 

4)  After the simulation of flow measured volume 

V (m

3

) value is read directly from the display 

(or through a digital interface) or calculated 

pulse counter calculates the volume pulse 

multiplied by the pulse value:

Permanent flow q

p

(m

3

/h)

Volume pulse value, 

(l/imp)

0.6

0.001

1.0

0.001

1.5

0.001

2.5

0.002

3.5

0.01

6.0

0.01

10.0

0.02

15.0

0.02

-  volume V (m

3

) measurement error calculated 

by formula:

V

 = (V – V

0

) / V

0

 • 100 % 

there:

V

0

 – device true value of simulated volume (m

3

).

3. Calibration instruction 

3.1.1 Calibration mode

To enter calibration mode it is necessary to 

connect the verification jumper 1, remove 

metrological seal and short circuit calibration 

contacts 2. (Fig.2).

3.2 Software installation

The software  “SonoMeter30Config.exe” is used 

to read and configure SonoMeter 30 device via 

OG-1-USB optical reader. 
Meter configuration / calibration cab be made 

also via M-Bus interface module then converter 

COM/M-Bus (MB-1) can be used.
Open “SonoMeter30Config.exe”. Press “Next”.

Then, you can choose installation directory.

“SonoMeter30UserConfig” has five user interface 

windows for different purposes:

•  “Actual” – actual data window;

•  “LCD Menu” – LCD configuration window. 

•  “Configuration” – configuration window for 

pulse channels, archive and tariffs;

3.3 Program interface

•  “Archive” – window for information from 

device archive;

•  “Manufacturer” – device calibration, reset and 

configuration window.

2.2 Volume measurement 

errors (δ

V

) test 

(continuous)

“SonoMeter30UserConfig” “Manufacturer” 

window is shown in Fig. 3, information about 

window parameters is in Table 1.

3.4 “Manufacturer” window

Fig. 2

1

2

3

4

5

6

7

8

9

10

Fig. 3
“Manufacturer” window