Dakewe(Shenzhen)Medical Equipment Co,. Ltd
43
V4.1
The section has chatter
lines, the section is
incomplete and breaks
midway
Blade is not fully locked
Lock the "blade lock handle";
Foreign matter is attached to
the blade installation groove,
which prevents the blade from
being fully clamped
Clean the knife surface of the
knife plate and the knife
surface of the tool holder seat
Tool holder seat is not stable
Lock the "knife holder seat
locking handle"
There is a rubbing sound
when slicing
The anti-roll plate is adjusted
too far forward, and the tissue
is scratched to the anti-roll
plate
Fine-tuning anti-roll plate
The section angle is too small,
and the tissue rubs against the
side of the rotating knife
holder
Adjust the rotating knife seat
to appropriately increase the
slice angle
The slice thickness is
uneven, even the knife
skips during the slicing
process
The sample holder is not fixed
Tighten the "sample tray
locking screw"
The trimming blade is too
large, the tissue is loose on the
sample, and the freezing is not
firm
Refreeze tissue with
embedding agent
The blade is loosen
Lock the "blade lock handle";
The blade holder is loosen
Lock the "blade holder seat
locking handle"
Damaged blade edge
Switch blade cutting edge
position or change blade
The material of the specimen
is uneven and too hard
The sample holder rotates 90
゜
and slices in another
direction
Slice angle too small
Adjust the rotating knife seat
to appropriately increase the
slice angle
Tissue pieces stick or
collapse on the anti-roll
board
Anti-roll plate temperature is
too high or the position is
incorrect
Lower the temperature and
continue to adjust the position
of the anti-roll plate
Anti-roll plate with static
electricity
Remove static electricity
Foreign matter on the edge of Clean the anti-roll plate and