7.3.2 Cleaning the Instrument Enclosure
The equipment shell is sprayed. Only soft cloth dampened with household cleaners, soapy
water, alcohol or alcohol-containing conventional disinfectants can be used for cleaning. Do
not use acetone, xylene or other corrosive solvents for cleaning and disinfection.
7.3.3 Cleaning the Blade Holder
Check if there is a blade, take the blade out first;
Loosen the right "blade lock handle" and pull it out, and then remove the knife plate;
Clean the surface of the pressure plate and the knife holder seat, focusing on cleaning the
surfaces where they are in contact with the blade;
Loosen the left "knife holder seat locking handle" and take out the entire knife holder seat
from the right;
Clean the tool holder seat and the rotating tool seat, focusing on cleaning the contact surface
between the two;
After cleaning, dry the parts and apply low-temperature lubricating oil on the contact
surface of the tool holder seat and the rotating tool seat;
Install and restore in order;
7.4 General Maintenance
Clean the inside of the machine freezer every day;
Lubricate the specimen feeding mechanism every week: press the fast forward button to
move the specimen chuck to the frontmost position, drop a drop of low-temperature
lubricating oil on the copper guide sleeve, and then press the fast reverse button to return
to the last position;
Lubricate the decorative board as needed;
Turn the handwheel, when the specimen chuck is at the highest position and the lowest
position, drop a few drops of low-temperature lubricant on the decorative guard, and then
turn the handwheel to make the lubricant evenly distributed;
Clean the condenser as needed
Open the condenser sealing baffle on the right side of the fuselage, and use a brush to clean
the dust on the condenser fins, or spray with high-pressure airflow.