6.8 If the test is started, the load is applied and released automatically. If the test is
completed, read the hardness value of the Digital Display part (①).
Measured Hardness
Mean Value
Measured Counts
Standard Deviation
User Setting
:
Measurement Setting Counts
Conversion to another hardness value: Convert into another
hardness value by using the Up/Down arrow keys (↑,↓).
6.9 Extract the specimen, turning the Screw Handle (⑥) to the right and continue the
next test. At this time, the Digital Display part (①) returns to the initial screen and
the Lighting Lamp (②) is turned on.
7. Cautions
7.1 Do not disassemble or modify this tester. For the repair,
contact our customer service team.
7.2 This tester operates on AC 220V. Do not touch with wet hands.
To do so may cause trouble or electirc shock.
7.3 If water enters in the tester, disconnect the power (AC 220V)
quickly. Contact your dealer or our customer service.
7.4 To transport the machine, first turn off the machine and then move. For
the installation procedure after transporting, refer to Paragraph 6.3..
7.5 For the test, process the contact surface roughly between a specimen and the Anvil to make it flat.