background image

18 

 

example above then it will dose the single add cap amount for each dose and then the last will be 
the remainder to make up the exact amount needed. 

Add KH interval: 

The gap between each automatic dose of KH buffer is determined by the ‘Add KH 

Interval’ setting, if the amount needed is greater than the single add cap. This setting can be used to 
spread dosing out between tests, it should be noted that enough time needs to be allowed between 
the last dose and the next test for the KH buffer to have mixed fully through the system and taken 
effect.  

For example, if using the information in the examples above, you are performing 4 tests a day (one 
test every 6 hours), the total add cap is 5ml and the single add cap is 1ml. From the information the 
maximum amount of times it could possibly dose is 5 times (1ml + 1ml + 1ml + 1ml + 1ml = 5ml), the 
goal will be to spread this out between the last test and the next test. The KH Manager will take 
approximately 25-30 minutes to perform a complete test cycle so there are 5 hours 30 minutes (330 
minutes) to complete the dosing as it is testing every 6 hours. If the add KH interval is set to 60 
minutes it can perform the 5 doses and still allow 30 minutes for the KH level to adjust before the 
next test (5 x 60 minutes = 300 minutes). 

To set the interval tap on ‘Add KH Interval’, then select the white box at the top of the page, enter 
the amount in minutes followed by ‘Done’, finally tap save in the top right corner to confirm and 
return to the settings page. 

KH Buffer Concentration:

 It is possible to use different brands of KH buffer solution with the KH 

Manager. For this to work correctly you need to program the concentration of the solution you are 
using into the machine. To do this tap on ‘KH Rise Reagent Brand’, in the next page there are two 
options ‘Kamoer’ and ‘Other’ if you are using Kamoer’s KH buffer then the strength is already 
programmed into the machine, simply tap on ‘Kamoer’ so that the blue tick is shown next to 
‘Kamoer’, then select ‘Save’ in the top right corner. If you are using another brand tap on ‘Other’ to 
select it and then enter the details of the buffer solution in the boxes below under the heading ‘Set 
Concentration of Reagent’. These details are usually displayed on the product bottle. To change each 
value, tap on the box in the app to bring up the keyboard, carefully enter the amount, then select 
done. Once the values are set select ‘Save’ in the top right corner to save and return to the settings 
page. 

Once all the settings are completed tap on the blue arrow in the top left corner to exit. 

 

SETTING CUSTOM PLAN MODE 

In this mode a user defined program is set up that the KH Manager follows, in the same way a 
programable dosing pump is used. However, if set, the machine can automatically vary the dose 
based on the test results to combat under or over dosing. This mode is useful if the KH consumption 
of the tank is constant and known. For example if you need to add 12ml of KH buffer solution to your 
aquarium each day, on average, to maintain a constant KH level you can set up a program telling the 
machine to add 0.5ml every hour throughout the day, if the automatic KH tests indicate the KH is 
starting to creep up then the machine will automatically decrease the 0.5ml dose, if the KH is 
starting to fall it will automatically increase the amount being dosed. The programming that the 
machine uses to work out the variance is user defined and can be done based on a percentage or 
volume.  

Содержание KH Manager

Страница 1: ...KH MANAGER INSTRUCTION MANUAL V1 0...

Страница 2: ...TO THE CLOUD 7 DIRECT BLUETOOTH CONNECTION TO THE KH MANAGER 7 STARTING THE KH MANAGER 8 IN APP STEP BY STEP GUIDE 8 INSTALING THE FRONT HOUSING 8 MIXING THE KH TEST REAGENT 9 CHECKING THE KH TEST REA...

Страница 3: ...R SOLUTION PUMP KH SOLUTION IN 23 CALIBRATING THE KH TEST PUMP KH REAGENT IN 24 CALIBRATING THE KH 25 RESETTING THE CALIBRATION ON THE KH MANAGER 25 MANUALLY RUNNING THE PUMP HEADS 26 RUNNING THE SALT...

Страница 4: ...ividually calibrated and easily serviced Data and programming stored on unit in case of power or connection loss Lab grade pH probe DIN port for connection and expansion with other D D Kamoer dosing p...

Страница 5: ...e contact your retailer immediately Box Contents Quick start guide KH Manager unit pH Probe Glass sample beaker 5ml Glass measuring cylinder 10ml Measuring cylinder High precision digital scales pH 4...

Страница 6: ...ht difference is within 0 5m The unit should be secured into place or installed in a stable location not above open water or in very humid environments The surface must be flat and level to ensure cor...

Страница 7: ...stream from where the sample water is being collected this will ensure a fresh sample is being taken for best accuracy when testing If the waste water is being sent to a waste liquid container then ne...

Страница 8: ...connection to a dedicated 5G Wi Fi network is not possible Once you have created an account log in and tap on the plus symbol in the middle of the screen if this is the first device being added or on...

Страница 9: ...rating specific components is described below section CALIBRATION the unit must be calibrated properly before use INSTALING THE FRONT HOUSING Once the KH Manager has been fully calibrated using the in...

Страница 10: ...eagent Supplied with the unit Weigh out exactly 980 grams of pure water RODI Add the complete 20ml bottle of standard concentrate reagent Gently mix the solution Mixing the double strength concentrate...

Страница 11: ...and then the blue arrow in the top left corner to exit the Maintain screen If the reagent is out of the acceptable range for use it should be safely discarded and a fresh solution made PERFORMING A T...

Страница 12: ...s displaying Automatic tap on the blue downwards arrow and select Manual test from the pop up at the bottom of the screen Next select the menu icon in the top right to enter the settings for the test...

Страница 13: ...tion will be added after the test as per the settings To turn off the automatic dosing tap on Tank Volume then tap on the toggle in the top right corner next to Keep KH to move it to the off position...

Страница 14: ...xit To turn the function off tap the toggle again and Off will be displayed to indicate it is not active Select the blue arrow in the top left corner to return to the test settings page As default the...

Страница 15: ...without losing any settings An immediate test function is also available when the KH Manager is in automatic mode This will allow a one off test without having to switch the machine to manual mode or...

Страница 16: ...tap on the Status icon in the bottom left corner A tile will be shown for the KH buffer solution KH Rise Reagent and the KH test reagent that displays the amount of liquid available the starting volum...

Страница 17: ...d the setting is shown in the separate KH Manager Expansion instructions There are two options that can be used Auto add and Custom Plans ACTIVATING AND DEACTIVATING AUTOMATIC KH DOSING NOTE The KH ri...

Страница 18: ...the pop up that appears on the bottom of the screen The settings on this page let you determine limitations on how much KH buffer solution can be added and how the machine doses the buffer solution T...

Страница 19: ...ou need to program the concentration of the solution you are using into the machine To do this tap on KH Rise Reagent Brand in the next page there are two options Kamoer and Other if you are using Kam...

Страница 20: ...hen enter a time for the plan to start by selecting Start Time use the scroll wheel to set the time and then save by selecting Done next tap on Add to enter the amount of solution you would like to ad...

Страница 21: ...ings page tap on the lower and upper parameters to make adjustments enter the figure then select Done and Save in the top right corner to save and exit Method This is used to determine if you would li...

Страница 22: ...ons some pure water RODI for rinsing the probe and some soft paper towel If present remove the protective cap from the pH probe by unscrewing the pot section and then sliding the seal and cap off the...

Страница 23: ...rom the KH Manager by pulling it forwards exposing the tubes When prepared tap I m ready in the app Place the glass beaker in the base of the unit under the short tube that is on the far right as you...

Страница 24: ...Calibration page will show you details of when the pump head was last calibrated when the next calibration is due this is a user defined setting the current head calibration data and the option to Go...

Страница 25: ...enter the number of days then save in the top right corner For accurate testing it is suggested to set this at 45 60 days after this time has elapsed a warning will appear on the screen in the app to...

Страница 26: ...sing Please recheck KH If this warning is shown perform a test with the KH Manager before commencing with this calibration Go to the calibration settings in the Kamoer Remote app and select KH from th...

Страница 27: ...for that pump to start the pump tap the corresponding box labelled Run the pump will start to add the amount specified and will show a countdown of that amount as it is dosed To stop mid dose tap the...

Страница 28: ...select Name from the list then tap on the white space and enter the new name for the unit When finished press Done followed by Save in the top right corner UPDATING THE KH MANAGER FIRMWARE Log into t...

Страница 29: ...If you want to reset the calibration settings on the KH Manager go to the main settings tab in the Kamoer Remote app select Maintain the calibration tile and then Reset Calibration tap on Reset and t...

Страница 30: ...ship be found within 12 months of the date of purchase D D The Aquarium Solution Ltd undertakes to repair or at our discretion replace the defective part free of charge Our policy is one of continual...

Отзывы: