The re-test function allows the KH Manager to perform a second test to confirm the result if the KH 
reading from the completed test varies too much from the last, this trigger point is user defined. For 
example, the last test gave a result of 7.5dKH and you have set a trigger point of 1.0dKH if the next 
test is above 8.5dKH or below 6.5dKH it will ignore the result and start another test to confirm if the 
reading was correct or not. If the end reading is still outside the trigger range the KH Manager will 
alert, if it is within the trigger range it will return to normal operation. To activate this function, 
select ‘Retest’ in the ‘Test Settings’ page then tap on the toggle in the top right to switch on. Set the 
trigger point by tapping on the box labelled ‘Trigger’ and enter the desired value. To turn the 
function off tap the toggle again and ‘Off’ will be displayed to indicate it is not active. Select the blue 
arrow in the top left corner to return to the test settings page. 
 
KH abnormality function can be set to increase the test frequency if the results fall outside the user 
defined threshold. For example, you have a KH threshold set of 7.2-7.8dKH and are testing every 12 
hours, if the test result gives a reading of 7.0dKH the KH Manager will increase the frequency of the 
tests to every 2 hours until the result is back above 7.2dKH. To set this function select ‘KH 
Abnormality Test’ in the ‘Test Settings’ page then tap on the toggle in the top right to switch on. Set 
the frequency to test by tapping on the box labelled ‘Test Interval’ and choose to re-test every 1,2,3 
or 4 hours then tap on the blue arrow in the top left to exit. To turn the function off tap the toggle 
again and ‘Off’ will be displayed to indicate it is not active. Select the blue arrow in the top left 
corner to return to the test settings page. 
 
As default the KH Manager is set with a test sample volume of 50ml, this volume is large enough to 
provide accurate results without using excessive KH test reagent per test cycle. It is however possible 
to increase the amount of sample water used up to a maximum of 100ml, using a larger sample 
volume will decrease the possible margin of error but can significantly increase the volume of 
reagent used and will create more waste water per test cycle. If increasing the test sample volume a 
larger 150ml glass test beaker will need to be obtained. To alter the volume tap on ‘Amount of Test 
Saltwater’ enter the preferred volume (can be between 50 and 100ml) then select save in the top 
right to confirm.  
 
Exit the test settings by tapping the blue arrow in the top right corner. 
 
The KH Manager has four pre-set times in automatic mode so the machine will perform a test every 
6 hours. These can be removed, added to, or adjusted if required.  
 

 

To adjust the time of a programmed test, tap on the time and then use the scroll wheel that 
appears to enter a new time, when done select ‘Save’ in the top right corner. 

 

For adding a new plan tap on the plus symbol towards the top right corner of the automatic 
test page, select the time of the new plan, and then choose ‘Save’ in the top right corner. 

 

Plans can be deleted by either swiping the plan to the left in the automatic test page and 
tapping the red delete box or by tapping on the plan and selecting ‘Delete’, then confirming 
by choosing ‘Delete’ in the pop-up that appears at the bottom of the screen. 

 

Please note a maximum of 24 tests can be set in a 24 hour period with a minimum gap of 60 
minutes between each test. 
 

Содержание KH Manager

Страница 1: ...KH MANAGER INSTRUCTION MANUAL V1 0...

Страница 2: ...TO THE CLOUD 7 DIRECT BLUETOOTH CONNECTION TO THE KH MANAGER 7 STARTING THE KH MANAGER 8 IN APP STEP BY STEP GUIDE 8 INSTALING THE FRONT HOUSING 8 MIXING THE KH TEST REAGENT 9 CHECKING THE KH TEST REA...

Страница 3: ...R SOLUTION PUMP KH SOLUTION IN 23 CALIBRATING THE KH TEST PUMP KH REAGENT IN 24 CALIBRATING THE KH 25 RESETTING THE CALIBRATION ON THE KH MANAGER 25 MANUALLY RUNNING THE PUMP HEADS 26 RUNNING THE SALT...

Страница 4: ...ividually calibrated and easily serviced Data and programming stored on unit in case of power or connection loss Lab grade pH probe DIN port for connection and expansion with other D D Kamoer dosing p...

Страница 5: ...e contact your retailer immediately Box Contents Quick start guide KH Manager unit pH Probe Glass sample beaker 5ml Glass measuring cylinder 10ml Measuring cylinder High precision digital scales pH 4...

Страница 6: ...ht difference is within 0 5m The unit should be secured into place or installed in a stable location not above open water or in very humid environments The surface must be flat and level to ensure cor...

Страница 7: ...stream from where the sample water is being collected this will ensure a fresh sample is being taken for best accuracy when testing If the waste water is being sent to a waste liquid container then ne...

Страница 8: ...connection to a dedicated 5G Wi Fi network is not possible Once you have created an account log in and tap on the plus symbol in the middle of the screen if this is the first device being added or on...

Страница 9: ...rating specific components is described below section CALIBRATION the unit must be calibrated properly before use INSTALING THE FRONT HOUSING Once the KH Manager has been fully calibrated using the in...

Страница 10: ...eagent Supplied with the unit Weigh out exactly 980 grams of pure water RODI Add the complete 20ml bottle of standard concentrate reagent Gently mix the solution Mixing the double strength concentrate...

Страница 11: ...and then the blue arrow in the top left corner to exit the Maintain screen If the reagent is out of the acceptable range for use it should be safely discarded and a fresh solution made PERFORMING A T...

Страница 12: ...s displaying Automatic tap on the blue downwards arrow and select Manual test from the pop up at the bottom of the screen Next select the menu icon in the top right to enter the settings for the test...

Страница 13: ...tion will be added after the test as per the settings To turn off the automatic dosing tap on Tank Volume then tap on the toggle in the top right corner next to Keep KH to move it to the off position...

Страница 14: ...xit To turn the function off tap the toggle again and Off will be displayed to indicate it is not active Select the blue arrow in the top left corner to return to the test settings page As default the...

Страница 15: ...without losing any settings An immediate test function is also available when the KH Manager is in automatic mode This will allow a one off test without having to switch the machine to manual mode or...

Страница 16: ...tap on the Status icon in the bottom left corner A tile will be shown for the KH buffer solution KH Rise Reagent and the KH test reagent that displays the amount of liquid available the starting volum...

Страница 17: ...d the setting is shown in the separate KH Manager Expansion instructions There are two options that can be used Auto add and Custom Plans ACTIVATING AND DEACTIVATING AUTOMATIC KH DOSING NOTE The KH ri...

Страница 18: ...the pop up that appears on the bottom of the screen The settings on this page let you determine limitations on how much KH buffer solution can be added and how the machine doses the buffer solution T...

Страница 19: ...ou need to program the concentration of the solution you are using into the machine To do this tap on KH Rise Reagent Brand in the next page there are two options Kamoer and Other if you are using Kam...

Страница 20: ...hen enter a time for the plan to start by selecting Start Time use the scroll wheel to set the time and then save by selecting Done next tap on Add to enter the amount of solution you would like to ad...

Страница 21: ...ings page tap on the lower and upper parameters to make adjustments enter the figure then select Done and Save in the top right corner to save and exit Method This is used to determine if you would li...

Страница 22: ...ons some pure water RODI for rinsing the probe and some soft paper towel If present remove the protective cap from the pH probe by unscrewing the pot section and then sliding the seal and cap off the...

Страница 23: ...rom the KH Manager by pulling it forwards exposing the tubes When prepared tap I m ready in the app Place the glass beaker in the base of the unit under the short tube that is on the far right as you...

Страница 24: ...Calibration page will show you details of when the pump head was last calibrated when the next calibration is due this is a user defined setting the current head calibration data and the option to Go...

Страница 25: ...enter the number of days then save in the top right corner For accurate testing it is suggested to set this at 45 60 days after this time has elapsed a warning will appear on the screen in the app to...

Страница 26: ...sing Please recheck KH If this warning is shown perform a test with the KH Manager before commencing with this calibration Go to the calibration settings in the Kamoer Remote app and select KH from th...

Страница 27: ...for that pump to start the pump tap the corresponding box labelled Run the pump will start to add the amount specified and will show a countdown of that amount as it is dosed To stop mid dose tap the...

Страница 28: ...select Name from the list then tap on the white space and enter the new name for the unit When finished press Done followed by Save in the top right corner UPDATING THE KH MANAGER FIRMWARE Log into t...

Страница 29: ...If you want to reset the calibration settings on the KH Manager go to the main settings tab in the Kamoer Remote app select Maintain the calibration tile and then Reset Calibration tap on Reset and t...

Страница 30: ...ship be found within 12 months of the date of purchase D D The Aquarium Solution Ltd undertakes to repair or at our discretion replace the defective part free of charge Our policy is one of continual...

Отзывы: