JF-1A Conductivity Sensor
A440-009
Page 19 of 56
FEBRUARY 2009
as calibration constants, or averaging settings, etc. The OPEN MODE can be
attained by sending the “***O” command. The unit after entering open mode
will send “OPEN MODE” when polled using a carriage return or line feed. From
open mode the user can view all constants using the “RCAL” command. The
RCAL Command will return the following:
***O
Open Mode
RCAL
1314
ZERO=55.000000
FS=1900.000000
BT=67.759354
MT=-9.510620E-05
TREF=20.100000
MC=1.280000E-02
N=1.000000
W=30
NREF=1.000000
DAO=375
DAF=500.000000
In the Table Below the definition of each of these constants is given:
Table 4
OPERATIONAL CONSTANTS
Constant
Definition
1314
Unit Serial Number
ZERO
The value assigned to the zero
reference reading, used to set
Y=MX+B for the conductivity
channel
FS
The value assigned to the Full
Scale reference reading, used to
set Y=MX+B for the conductivity
channel
BT
The
Temperature
Channel
Offset
Value
MT
The
Temperature
Channel
Slope
Value
TREF
The
Temperature
Reference
for
Conductivity Compensation
MC
The Temperature Coefficient of Jet