System Check or Laser Cal Condition Table
DYE, Sys Check at 750 Volts Eresonator
DYE Eres
Joules
Cap
Voltage
Fiber HP XSM
Condition and Message
> 2.5
750
> 65%
Good resonator cal data with 20%
> 2.5
750
> 120%
E33: DYE Calibration Error Call Service
> 2.5
750
50 to 65%
W59: Low Trans. Check Fiber/Handpiece
> 2.5
750
< 50%
E34: DYE Cal Failed Energy Low Call Service
< 2.5
750
E34: DYE Cal Failed Energy Low Call Service
> 2.5
750
Cal Port <1J
E26: No energy check fiber/handpiece
DYE, Laser Cal, valid for 0.5 ms
DYE Eres
Joules
Cap
Voltage
Fiber HP XSM
Condition and Message
> 5
< 1000
Good
< 5
>= 1000
W73:Schedule preventive maintenance
< 4
>= 1000
E34: DYE Cal Failed Low Energy Call Service
> 12
W67: Unable to Reach Fluence; Decrease
> 5
>= 1000
W67: Unable to Reach Fluence; Decrease
< 50%
E23: Low Trans. Replace Fiber/Handpiece
Nd:YAG, Sys Check at 750 Volts Eresonator
DYE Eres
Joules
Cap
Voltage
Fiber HP XSM
Condition and Message
> 30
750
> 65%
Good resonator cal data with 20%
> 30
750
> 120%
E24: YAG Calibration Error Call Service
> 30
750
50% to 65%
W59: Low Trans. Check Fiber/Handpiece
> 30
750
< 50%
E25:YAG Cal Failed Energy Low Call Service
< 30
750
E25:YAG Cal Failed Energy Low Call Service
> 30
750
Cal Port <5J
E26: No energy check fiber/handpiece
Nd:YAG, Laser Cal, valid for 5 and 10 ms
DYE Eres
Joules
Cap
Voltage
Fiber HP XSM
Condition and Message
> 60
< 1000
Good
< 60
>= 1000
W69: Schedule Lamp Change
< 50
>= 1000
E25: YAG Cal Failed Energy Low Call Service
> 95
W67: Unable to Reach Fluence; Decrease
> 60
>= 1150
W67: Unable to Reach Fluence; Decrease
< 50%
E23: Low Trans. Replace Fiber/Handpiece
Sys Check at 750 Volts Eresonator Combination
DYE Eres
Joules
Cap
Voltage
Fiber HP XSM
Dye HP XSM
Condition and Message
< 30
< 2.5
E40: DYE and YAG Cal Failed, Call Service
> 30
> 2.5
< 50%
< 50%
E23: Low Trans. Replace Fiber/Handpiece
> 30
> 2.5
< 50%
< 65%
E23: Low Trans. Replace Fiber/Handpiece
> 30
> 2.5
< 65%
< 50%
E23: Low Trans. Replace Fiber/Handpiece
Содержание Cynergy
Страница 1: ...Cynergy 850 1265 000 Rev 4 TECHNICAL GUIDE CYNOSURE Where art and science meet ...
Страница 10: ...8 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Страница 14: ...12 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Страница 18: ...16 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Страница 22: ...20 of 112 850 1265 000 Rev 4 Cynosure Inc Figure 2A Performance Diagnostic Flow Diagram ...
Страница 23: ...Cynergy Technical Guide 850 1265 000 Rev 4 21 of 112 Figure 2B Normal Operation Flow Diagram ...
Страница 24: ...22 of 112 850 1265 000 Rev 4 Cynosure Inc Figure 2C Cynergy MultiPlex Flow Diagram ...
Страница 38: ...36 of 112 850 1265 000 Rev 4 Cynosure Inc Figure 12B Frame w Electrical Module Rear View ...
Страница 43: ...Cynergy Technical Guide 850 1265 000 Rev 4 41 of 112 Figure 14B Fiber Block Beam Path ...
Страница 46: ...44 of 112 850 1265 000 Rev 4 Cynosure Inc Figure 16 Dye Tank ...
Страница 47: ...Cynergy Technical Guide 850 1265 000 Rev 4 45 of 112 Figure 17 Dye Concentration Sensor Figure 18 Dye Pump ...
Страница 52: ...50 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Страница 64: ...62 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Страница 67: ...Cynergy Technical Guide 850 1265 000 Rev 4 65 of 112 Figure 24 Dye Injection Diagnostic Routine Flowchart ...
Страница 82: ...80 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Страница 84: ...82 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Страница 86: ...84 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Страница 90: ...88 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Страница 98: ...96 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Страница 106: ...104 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Страница 114: ...112 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Страница 115: ...Tabbed Page Final Test Calibration ...
Страница 116: ......
Страница 118: ......
Страница 120: ......
Страница 148: ......
Страница 149: ...Tabbed Page Drawing Sets ...
Страница 150: ......
Страница 172: ......
Страница 173: ...Tabbed Page Schematics ...
Страница 174: ......
Страница 176: ...This page intentionally left blank ...
Страница 178: ...This page intentionally left blank ...
Страница 180: ...This page intentionally left blank ...
Страница 182: ...This page intentionally left blank ...
Страница 184: ...This page intentionally left blank ...
Страница 186: ...This page intentionally left blank ...
Страница 188: ...This page intentionally left blank ...
Страница 190: ...This page intentionally left blank ...
Страница 192: ...This page intentionally left blank ...
Страница 194: ...This page intentionally left blank ...