Cynergy Final Test and Calibration Procedure Doc #991-4906-000
Page 25 of 28
Rev 5
Appendix A – CYNERGY Error Codes
Cynergy Error and Warning Codes, October 6, 2005
Possible Cause
Error #
Message Text
Manual Suggestions
E01
Water Flow Fault; Cycle Power
Broken water pump or blown fuse. Needs
service.
E02
Open Shutter Fault
Safety shutter stuck closed. Do not use
laser until serviced.
E03
Close Shutter Fault
Safety shutter stuck open. Do not use laser
until serviced.
E04
Fuse Open; Call Service
Blown capacitor bank fuse or shorted dump
circuit. Needs service.
E05
Checksum Error; Call Service
Computer program memory error. Needs
service.
E07
IGBT Fault - YAG
YAG IGBT overload sensed. Cycle power to
continue.
E08
Heater Fault; Call Service
Open thermostat or heater. Blown heater
fuse. Needs service.
E09
Coolant Sensor Fault; Call Service
Open thermistor sensor or unplugged.
Needs service.
E10
No High Voltage
Capacitor bank or HVPS wire shorted. No
read back voltage. Needs service.
E13
No Simmer YAG
Broken lamp or blown fuse. Needs service.
E14
HVPS Fault
HVPS over-temperature.
E15
Dump Fault; Call Service
Fuse fault at power up. Needs service.
E16
Dump Fault; Call Service
Dump always on. Needs service. Dump
shorted. Failed dump TX. HVPS failed.
E17
Dump Fault; Call Service
Dump not working. Needs service. Fiber
connections broken to cap bank. HVPS
disconnected or failed.
E23
Low Trans; Replace Fiber/Handpiece
Damaged fiber, handpiece or focus lens,
needs replacement.
E24
YAG Calibration Error; Call Service
Incorrect Resonator Port or Cal Port sensor
calibration. Needs service.
E25
YAG Cal Failed, Energy Low; Call Service
YAG cal port energy too low to use. Needs
service.
E26
No Energy; Check Fiber/Handpiece
Broken fiber or severly damaged focus lens.
Needs replacement.
E27
DYE Flow Fault
No dye filter or it is not installed properly.
Broken pump. Blown fuse
E28
IGBT Fault - Dye
Dye IGBT overload sensed. Cycle power to
continue.
E29
IGBT Fault - Pulse Simmer
Pulse simmer IGBT overload sensed. Cycle
power to continue.
E30
No Simmer DYE
Broken lamp or blown fuse. Needs service.
Содержание Cynergy
Страница 1: ...Cynergy 850 1265 000 Rev 4 TECHNICAL GUIDE CYNOSURE Where art and science meet ...
Страница 10: ...8 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Страница 14: ...12 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Страница 18: ...16 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Страница 22: ...20 of 112 850 1265 000 Rev 4 Cynosure Inc Figure 2A Performance Diagnostic Flow Diagram ...
Страница 23: ...Cynergy Technical Guide 850 1265 000 Rev 4 21 of 112 Figure 2B Normal Operation Flow Diagram ...
Страница 24: ...22 of 112 850 1265 000 Rev 4 Cynosure Inc Figure 2C Cynergy MultiPlex Flow Diagram ...
Страница 38: ...36 of 112 850 1265 000 Rev 4 Cynosure Inc Figure 12B Frame w Electrical Module Rear View ...
Страница 43: ...Cynergy Technical Guide 850 1265 000 Rev 4 41 of 112 Figure 14B Fiber Block Beam Path ...
Страница 46: ...44 of 112 850 1265 000 Rev 4 Cynosure Inc Figure 16 Dye Tank ...
Страница 47: ...Cynergy Technical Guide 850 1265 000 Rev 4 45 of 112 Figure 17 Dye Concentration Sensor Figure 18 Dye Pump ...
Страница 52: ...50 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Страница 64: ...62 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Страница 67: ...Cynergy Technical Guide 850 1265 000 Rev 4 65 of 112 Figure 24 Dye Injection Diagnostic Routine Flowchart ...
Страница 82: ...80 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Страница 84: ...82 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Страница 86: ...84 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Страница 90: ...88 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Страница 98: ...96 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Страница 106: ...104 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Страница 114: ...112 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Страница 115: ...Tabbed Page Final Test Calibration ...
Страница 116: ......
Страница 118: ......
Страница 120: ......
Страница 148: ......
Страница 149: ...Tabbed Page Drawing Sets ...
Страница 150: ......
Страница 172: ......
Страница 173: ...Tabbed Page Schematics ...
Страница 174: ......
Страница 176: ...This page intentionally left blank ...
Страница 178: ...This page intentionally left blank ...
Страница 180: ...This page intentionally left blank ...
Страница 182: ...This page intentionally left blank ...
Страница 184: ...This page intentionally left blank ...
Страница 186: ...This page intentionally left blank ...
Страница 188: ...This page intentionally left blank ...
Страница 190: ...This page intentionally left blank ...
Страница 192: ...This page intentionally left blank ...
Страница 194: ...This page intentionally left blank ...