![Cynosure Cynergy Скачать руководство пользователя страница 109](http://html.mh-extra.com/html/cynosure/cynergy/cynergy_technical-manual_2704655109.webp)
Cynergy Technical Guide
850-1265-000, Rev. 4
107 of 112
Fault Code and Text Message
Probable Cause/Corrective Action
W58 Handpiece
Changed
Handpiece changed. Press screen message box to
continue.
W59
Low Trans; Check Fiber and
Handpiece
Fiber, handpiece, or focus lens damaged but still
useable.
W60
YAG Cal Failed, Energy Low; Call
Service
Low delivered yag energy. Needs service
W61
PDL Cal Failed, Energy Low; Call
Service
Low delivered dye energy. Needs service
W63
Fluence Meter Error; Call Service
Miscalibrated cal port meter. Defective cal port. Needs
service.
W64
Internal Meter Error; Call Service
Miscalibrated res port meter. Defective res port. Needs
service.
W65
Energy Out of Range, YAG
Capacitor bank did not recharge properly or
overcharged. Retry operation.
W66 Handpiece
Removed
Handpiece removed from cal port during system check
or calibration.
W67
Unable to Reach Fluence;
Decrease
Max fluence reached for handpiece and/or treatment
settings. Lamps near end-of life.
W68
Schedule Lamp Change, PDL
Lamps for dye laser approaching end-of-life. Schedule
lamp replacement.
W69
Schedule Lamp Change, YAG
Lamps for Nd:YAG laser approaching end-of-life.
Schedule lamp replacement.
W70
Coolant Over Temp; Let Laser
Cool
Water is too hot for continued operation. Fans are on to
cool laser. Let laser cool.
W71
Add Water Soon
Water level low - refill water reservoir.
W72
Energy Out of Range, PDL
Capacitor bank did not recharge properly or
overcharged. Retry operation.
W73
PDL Over Temp; Let Laser Cool
Dye is too hot for continued operation. Fans are on to
cool laser. Let laser cool.
W74
Warning: Purge Dye Concentrate
Bottle
Air in dye line, squeeze bottle
W75
Dye Mix Warning; Call Service
<ABS 0.85 after 30 min cycle (30 injects)
Self Test
During startup, the laser runs a self test. If the fault occurs during this process, consult the fault
code chart above for information on resolving the fault.
Содержание Cynergy
Страница 1: ...Cynergy 850 1265 000 Rev 4 TECHNICAL GUIDE CYNOSURE Where art and science meet ...
Страница 10: ...8 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Страница 14: ...12 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Страница 18: ...16 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Страница 22: ...20 of 112 850 1265 000 Rev 4 Cynosure Inc Figure 2A Performance Diagnostic Flow Diagram ...
Страница 23: ...Cynergy Technical Guide 850 1265 000 Rev 4 21 of 112 Figure 2B Normal Operation Flow Diagram ...
Страница 24: ...22 of 112 850 1265 000 Rev 4 Cynosure Inc Figure 2C Cynergy MultiPlex Flow Diagram ...
Страница 38: ...36 of 112 850 1265 000 Rev 4 Cynosure Inc Figure 12B Frame w Electrical Module Rear View ...
Страница 43: ...Cynergy Technical Guide 850 1265 000 Rev 4 41 of 112 Figure 14B Fiber Block Beam Path ...
Страница 46: ...44 of 112 850 1265 000 Rev 4 Cynosure Inc Figure 16 Dye Tank ...
Страница 47: ...Cynergy Technical Guide 850 1265 000 Rev 4 45 of 112 Figure 17 Dye Concentration Sensor Figure 18 Dye Pump ...
Страница 52: ...50 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Страница 64: ...62 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Страница 67: ...Cynergy Technical Guide 850 1265 000 Rev 4 65 of 112 Figure 24 Dye Injection Diagnostic Routine Flowchart ...
Страница 82: ...80 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Страница 84: ...82 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Страница 86: ...84 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Страница 90: ...88 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Страница 98: ...96 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Страница 106: ...104 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Страница 114: ...112 of 112 850 1265 000 Rev 4 Cynosure Inc This page has been intentionally left blank ...
Страница 115: ...Tabbed Page Final Test Calibration ...
Страница 116: ......
Страница 118: ......
Страница 120: ......
Страница 148: ......
Страница 149: ...Tabbed Page Drawing Sets ...
Страница 150: ......
Страница 172: ......
Страница 173: ...Tabbed Page Schematics ...
Страница 174: ......
Страница 176: ...This page intentionally left blank ...
Страница 178: ...This page intentionally left blank ...
Страница 180: ...This page intentionally left blank ...
Страница 182: ...This page intentionally left blank ...
Страница 184: ...This page intentionally left blank ...
Страница 186: ...This page intentionally left blank ...
Страница 188: ...This page intentionally left blank ...
Страница 190: ...This page intentionally left blank ...
Страница 192: ...This page intentionally left blank ...
Страница 194: ...This page intentionally left blank ...