MHW 0008 – DK 1695  |  Version 002 – 21/08/2019  |  custo med GmbH

2

Holter

 Hardware, description of device for custo watch

[2] Holter hardware, custo watch | 

page 7

2.4 

Charging custo watch and custo guard

Connect the custo docking station with the micro USB cable   to the switched-on 

PC. Only use the supplied micro USB cable. Alternatively, the custo docking station 

can be connected to the power supply with the plug-in power supply unit.

Place the custo watch onto the custo docking station. The custo watch must engage 

noticeably and be firmly seated on the custo docking station, with the grey button 

pointing downwards  . The custo guard 3 ECG transmitter is also charged in the 

custo docking station. The direction of insertion is determined by the shape of the 

housing (the opening of the custo med logo points upwards 

  ).

Charging- & running times 

custo watch 

custo guard 3

max. charging time 

approx. 3h 

approx. 3 h

max. running time 

1)

 

approx. 4 days 

approx. 3 days

Sampling rate 

--- 

1 kHz

Standby 

approx. 30 days 

approx. 100 days

Display of operational readiness 

2)

, rechargeable battery capacity

The LED on the custo guard 3 ECG transmitter   indicates the operational readiness 

of the system when both devices are in the custo docking station.

LED RED

 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 

Battery capacity of at least one device < 24 hours

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 

and/or recording not yet downloaded

LED GREEN

 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Battery capacity of both devices > 24 hours

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

and recording has been downloaded = ready for recording

If the grey button on the custo watch is pressed   while both devices are in the custo 

docking station, the rechargeable battery capacity of both devices is shown on the 

display. If there is sufficient rechargeable battery capacity for a 24-hour recording 

period, a checkmark appears for the corresponding device. If the rechargeable bat-

tery capacity is less than 24 hours, the current charge status is displayed. If there is 

no custo guard 3 ECG transmitter on the custo docking station or an incorrect custo 

guard 3 (i.e. one which is not connected to the custo watch), the display shows 

guard: ---. Place the correct custo guard 3 onto the custo docking station.

Frequency of charging cycles

We recommend charging the devices after each recording. Clean, disinfect and dry 

the devices before recharging.

1)  The rechargeable battery 

runtime is dependent, among 
other things, on the distance and 
obstacles between the receiver 
and transmitter (condition of the 
transmission path) and the age 
and condition of the rechargeable 
battery. After 300 or more charging 
cycles, the battery has a capacity of 
80 % of the initial capacity.

Defi nition of charge cycle:
During a charge cycle the entire 
battery power is used, however, this 
does not have to occur in a single 
charging operation. For example, 
you can use a device for several 
hours on a particular day, use 
approximately half the capacity, 
and then fully charge it again. If 
the same procedure is repeated on 
the following day, this is equal to 
just one charging cycle and not two 
charging cycles. It can therefore 
take a few days for a charge cycle to 
be fully completed.

2) Prerequisite: 

custo watch and custo guard have 
already been connected with each 
other in custo diagnostic and the 
serial number of the custo guard 
ECG transmitter is stored in the 
custo watch. The devices are 
connected during the start 
procedure (refer to Part 3, Holter - 
description of software, chapter 

“Performing a holter recording”).

It is only possible to display 
operational readiness if no recording 
is currently in progress and both 
devices are in the custo docking 
station that is connected to the PC.

IMPORTANT:

Charge 

the devices 

before 

first use!

Содержание custo belt

Страница 1: ...Manual Holter Part 2 Hardware description of device for custo watch 1 2 3 4 Safety Hardware Software Hygiene Operating characteristics For 3 channel holter up to 72 hours MHW 0008 DK 1695 Version 002...

Страница 2: ...nother language without the prior written consent of custo med GmbH The manufacturer reserves the right to change the information in this Operating Manual without prior notice The current version can...

Страница 3: ...play 10 2 6 Attaching the holter recorder to the patient 2 6 1 Fitting the custo watch 13 2 6 2 Fitting the custo guard ECG transmitter with electrode belt 13 2 6 3 Length adjustment of the custo belt...

Страница 4: ...r custo med GmbH Maria Merian Str 6 85521 Ottobrunn Germany CE mark Medical electrical equipment classification according to DIN EN 60601 1 Type BF Observe the Operating Manual Protection class of ele...

Страница 5: ...perfectly safe for patients with a pacemaker Pacemaker recognition is not possible with the device Caution The ECG recording may be affected by pacemaker pulses The system is intended for use by trai...

Страница 6: ...usto guard 3 3 channel ECG transmitter custo belt 3 red 3 channel electrode belt custo belt 1 3 extender extension piece for the electrode belt Washing net for custo belt custo docking station charger...

Страница 7: ...mark appears for the corresponding device If the rechargeable bat tery capacity is less than 24 hours the current charge status is displayed If there is no custo guard 3 ECG transmitter on the custo d...

Страница 8: ...dicates the opera tional readiness of the system when both devices are in the custo docking station and connected to the PC or power supply unit 2 5 2 Changing the custo watch strap set To change the...

Страница 9: ...ppears when the connection to the ECG transmitter is successful ECG symbol appears when QRS complexes are detected Recording symbol appears when the ECG recording has been started Duration of current...

Страница 10: ...ay by briefly pressing the grey button To ensure a full 24 hour or 72 hour recording both devices must be sufficiently charged If the custo watch is outside the custo docking station the display is sw...

Страница 11: ...played see The grey button is then held pressed approx 10 seconds until the Finish recording query appears in the display The recording is ended by briefly pressing the grey button again The message R...

Страница 12: ...ge 12 Memory full The memory card in the custo watch is full The custo watch must be downloaded into custo diagnostic before a new recording can be performed The data are stored in the custo diagnosti...

Страница 13: ...oiled Soiling will affect the quality of the ECG1 CAUTION The electrode contact surfaces must always be in contact with the skin Ensure that there are no textiles between the electrode and the skin Ma...

Страница 14: ...ed directly next to the point where the material becomes wider on the custo belt The elastic excess length is completely pulled through the slider so that the slider is directly next to the black plas...

Страница 15: ...k and the wide material point Extension with the custo belt extender If the maximum length is not sufficient use a custo belt extender extension piece The custo belt extender is adjusted to the minimu...

Страница 16: ...sensive to the cable ends Attach the neck strap to the hooks of the ECG cable custo guard 3 Put the ECG transmitter around the patients neck and adjust the neck strap to the desired length Clean the...

Страница 17: ...er A custo belt electrode belt stretches after prolonged wear due to moisture sweat The custo belt should be re tensioned after some time so that it rests firmly against the upper body again and to en...

Страница 18: ...Pa Classification Device with internal power supply Class IIa Type BF Underlying standards DIN EN 60601 1 DIN EN 60601 1 2 DIN IEC 60601 2 47 Technical data custo docking station Connection to the PC...

Страница 19: ...10 95 rH Air pressure 700 1060 hPa Classification Device with internal power supply Class IIa Type BF Underlying standards DIN EN 60601 1 DIN EN 60601 1 2 DIN EN 60601 2 47 Technical data custo belt 3...

Страница 20: ...r 2012 R older versions are not supported PC The PC hardware should meet the minimum requirements of the operating system used Provide additional RAM 1 GB for custo diagnostic Please ensure that there...

Страница 21: ...sition systems Fanless Dual DVI or DP graphics card for CPET Windows 8 1 x64 PRO version for joining a domain Ports One USB 2 0 port per USB device preferably not USB 3 0 One COM port each for ergomet...

Страница 22: ...ccording to IEC61000 3 2 Voltage fluctuations flickers according to IEC61000 3 3 Compliance Group 1 Class B Not applicable Not applicable Electromagnetic environment guidelines custo watch uses the fr...

Страница 23: ...80 MHz are 6 765 to 6 795 MHz 13 553 MHz to 13 567 MHz 26 957 MHz to 27 283 MHz and 40 66 MHz to 40 70 MHz The amateur radio bands between 0 15 MHz and 80 MHz are 1 8 MHz to 2 0 MHz 3 5 MHz to 4 0 MH...

Страница 24: ...ided as this could result in fault operation If the devices must nonetheless be used as described above this device and the other devices should be monitored to ensure proper functionality a For some...

Страница 25: ...MHW 0008 DK 1695 Version 002 21 08 2019 custo med GmbH 2 Holter Hardware description of device for custo watch 2 Holter hardware custo watch page 25 2 10 EC Declaration of Conformity...

Страница 26: ...MHW 0008 DK 1695 Version 002 21 08 2019 custo med GmbH 2 Holter Hardware description of device for custo watch 2 Holter hardware custo watch page 26...

Страница 27: ...Z 1 Strap set for custo watch large circumference 180 260 mm 10302 1 custo guard 3 10301 R 1 custo belt 3 red for custo guard 3 10118 1 custo belt 1 3 extender for custo guard 1 3 10121 1 1 ECG cable...

Страница 28: ...Manufacturer s contact details custo med GmbH Maria Merian Stra e 6 85521 Ottobrunn Germany Phone 49 0 89 710 98 00 Fax 49 0 89 710 98 10 E mail info customed de Internet www customed de...

Отзывы: