![Custo Med custo belt Скачать руководство пользователя страница 22](http://html1.mh-extra.com/html/custo-med/custo-belt/custo-belt_operating-manual_2700571022.webp)
MHW 0008 – DK 1695 | Version 002 – 21/08/2019 | custo med GmbH
2
Holter
Hardware, description of device for custo watch
[2] Holter hardware, custo watch |
page 22
2.9
Manufacturer’s declaration regarding EMC (electromagnetic compatibility)
according to DIN EN 60601-1-2:2016-05
Cable lengths:
ECG cable custo guard 3 & cable adapter: 175 mm, micro USB cable 2.0: 1,5 m
Manufacturer's declaration – electromagnetic emissions
The custo watch holter recorder (consisting of custo watch and ECG transmitter)) is designed for use in the elec-
tromagnetic environment stated below. The customer or user of the custo watch holter recorder should make
sure that it is used in such an environment.
Manufacturer’s declaration – electromagnetic immunity
The custo watch holter recorder (consisting of custo watch and ECG transmitter)) is designed for use in the elec-
tromagnetic environment stated below. The customer or user of the custo watch holter recorder should make
sure that it is used in such an environment.
COMMENT: U
T
is the alternating supply voltage prior to application of test levels
Emission measurements
RF emissions according to CISPR11
RF emissions according to CISPR11
Harmonics according to IEC61000-3-2
Voltage fluctuations/flickers according to IEC61000-3-3
Compliance
Group 1
Class B
Not applicable
Not applicable
Electromagnetic environment - guidelines
custo watch uses the frequency band in the range 2.4
GHz for communication with the custo guard ECG trans-
mitter. RF emissions are very low and it is unlikely that
adjacent electronic devices will be disturbed.
custo watch is designed for use in all establishments, in-
cluding domestic establishments and those directly con-
nected to the public low voltage power supply network
that supplies buildings used for domestic purposes.
Immunity tests
Static electricity discharge (ESD)
according to IEC 61000-4-2
Quick transient electric interference factors / bursts
according to IEC 61000-4-4
Surges
according to IEC 61000-4-5
Voltage drops, brief interruptions
and fluctuations in supply voltage
according to IEC 61000-4-11
Magnetic field at supply frequency (50/60 Hz)
according to IEC 61000-4-8
IEC 60601- test level
± 8 kV contact discharge
± 15 kV air discharge
± 2 kV for net wires
± 1 kV for input and
Output leads (SIP/SOP)
± 1 kV lead against lead
± 2 kV lead against end
< 5 % U
T
for 0.5 periods
(> 95 % drop)
40 % U
T
for 5 periods
(60 % drop)
70 % U
T
for 25 periods
(30 % drop)
< 5 % U
T
for 5 s
(> 95 % drop)
30 A/m
Compliance level
± 8 kV contact discharge
± 15 kV air discharge
Not applicable
Not applicable
Not applicable
30 A/m