42
Culligan® Premier Series Automatic Deionizer
42
Cat. No. 01024000
Diagnostics
Manual/Motor Control
Use the deionizer controller to manually control the duration of the deionization system processes.
SERVICE
FEB-01-13 12:11P
10.0GPM 161PPM
1. From the default home screen, press
to display the main menu.
3)SET DATE/TIME
4)ADV. SETUP
5)ACCESSORIES
>6)DIAGNOSTICS
2. Press
to select 6) DIAGNOSTICS, the diagnostics informa-
tion and settings.
>1)MANUAL CONTROL
2)CHECK INPUTS
3)TEST PHONELIN
4)TEST PROGFLOW
3. Press
to select 1) MANUAL CONTROL.
BACKWASH CAT
MINUTES LEFT:6
>NEXT CYCLE
4. If the system is regenerating, the first screen displayed is the remaining minutes
for cation backwash (BACKWASH CAT).
5. Press
.
The screen displays a prompt to view the next cycle.
SLOW RINSE CAT
MINUTES LEFT:30
>END REGEN NOW
6. The next diagnostic screen displays the current state of the deionizer tank in ser-
vice (such as SLOW RINSE CAT).
7. Press
.
The screen displays a prompt to end regeneration mode.
DEIONIZER 1 IN
SERVICE
8. Press
.
The first diagnostic screen displays the current state of the deionizer
tank in service.
SERVICE
FEB-01-13 12:11P
10.0GPM 161PPM
9. The controller automatically returns to the home screen after a few moments.