10 Troubleshooting
10.1 Self-Tests
The
Paramedic CU-ER1
executes automated and manual self-tests to verify the
functionality of its major subsystems.
The
Paramedic CU-ER1
raises alarms if it detects a fault during those self-tests. It raises
more alarms when it is turned ON after a fault is detected.
Alarms
Alarms raised after a failure is detected during Power On or Run Time Test:
1.
The
Paramedic CU-ER1
displays the error name and code on the LCD screen
2.
The ERROR indicator light turns ON.
3.
The beeper emits short beeps at regular intervals
Alarms raised after a fault is detected during self-tests (automatic and manual):
1.
The ERROR indicator light turns ON then OFF.
2.
The beeper emits a single beep.
3.
The alarms stated in 1 and 2 are repeated every minute while the Paramedic is
OFF.
Alarms raised when the
Paramedic CU-ER1
is turned ON after a fault is detected during
automatic or manual self-test:
1.
The
Paramedic CU-ER1
displays the error name and code on the LCD screen
2.
The ERROR indicator light turns ON.
3.
The beeper emits short beeps at regular intervals
Self-Test Errors
Error Name
Code
Fault
Remedial Action
LOW BATTERY
0001
Low battery charge
Recharge the battery
SYSTEM ERROR
0002
NAND Flash memory failure
Return to an authorized
service center for repair
SYSTEM ERROR
0004
Random Access Memory
(RAM) Failure
Return to an authorized
service center for repair
SYSTEM ERROR
0008
Real Time Clock (RTC)
Failure
Return to an authorized
service center for repair
SYSTEM ERROR
0016
Complex Programmable
Logic Device
Return to an authorized
service center for repair
76
Содержание Paramedic CU-ER1
Страница 1: ...CU S Emergency Responder 1 Operator s Manual 1 CU Medical Systems Inc Paramedic Series AED CU ER1 ...
Страница 2: ...CU ER1 Operator s Manual _ver 3 00 Paramedic CU ER1 Quick Reference Card 2 ...
Страница 90: ...CU ER1 Operator s Manual _ver 3 00 Notes 90 ...
Страница 113: ...CU ER1 Operator s Manual _ver 3 00 75Ω Impedance Load 150 Joules output 100Ω Impedance Load 150 Joules output 113 ...
Страница 114: ...CU ER1 Operator s Manual _ver 3 00 125Ω Impedance Load 150 Joules output 150Ω Impedance Load 150 Joules output 114 ...
Страница 115: ...CU ER1 Operator s Manual _ver 3 00 175Ω Impedance Load 150 Joules output 115 ...