CU-ER1 Operator’s Manual _ver 3.00
4.4 Voice and Text Prompts
Prompt Type/Meaning
ATTACH PADS
Voice and Text
Indicates that you have to attach the defibrillator electrode
pads to the bare chest wall of the patient.
DO NOT TOUCH THE
PATIENT
Voice and Text
Indicates that the Paramedic CU-ER1 is analyzing the ECG
signal acquired from the patient. You must not touch the
patient to minimize artifacts. Motion during ECG
acquisition introduces artifacts.
ANALYZING HEART RHYTHM
Voice and Text
Indicates that the device is doing an analysis of the
patient’s ECG.
SHOCK ADVISED
Voice and Text
Indicates that the patient has a shockable ECG rhythm.
STAND CLEAR
Voice and Text
Indicates that everybody in the vicinity of the patient must
stand clear and not touch the patient.
CHARGING
Text
Indicates that the Paramedic CU-ER1 is charging its
capacitor to prepare for a shock delivery.
CHARGING COMPLETE
Text
Indicates that the Paramedic CU-ER1 has finished charging
its capacitor. This prompt is in text form only.
PRESS THE FLASHING RED
BUTTON, NOW
Voice and Text
Indicates that you have to press the SHOCK button for the
delivery of a defibrillation shock. At this time, the SHOCK
button is flashing and the beeper is beeping.
SHOCK DELIVERED
Voice and Text
Indicates that the Paramedic CU-ER1 has delivered a
defibrillation shock.
28
Содержание Paramedic CU-ER1
Страница 1: ...CU S Emergency Responder 1 Operator s Manual 1 CU Medical Systems Inc Paramedic Series AED CU ER1 ...
Страница 2: ...CU ER1 Operator s Manual _ver 3 00 Paramedic CU ER1 Quick Reference Card 2 ...
Страница 90: ...CU ER1 Operator s Manual _ver 3 00 Notes 90 ...
Страница 113: ...CU ER1 Operator s Manual _ver 3 00 75Ω Impedance Load 150 Joules output 100Ω Impedance Load 150 Joules output 113 ...
Страница 114: ...CU ER1 Operator s Manual _ver 3 00 125Ω Impedance Load 150 Joules output 150Ω Impedance Load 150 Joules output 114 ...
Страница 115: ...CU ER1 Operator s Manual _ver 3 00 175Ω Impedance Load 150 Joules output 115 ...