10
GG-H2S
Specifications
Input Power:
+24 VDC, 350 mA
Detection Principle:
Electrochemical
Detection Method:
Diffusion
Gases:
Hydrogen sulfide (H2S)
Ranges:
0/50 ppm
Output Signal:
Linear 4/20 mA (max input impedance: 700 Ohms)
Response Time:
T
50
= less than 30 seconds
T
90
= less than 120 seconds
Accuracy:
+/- 5% of value, but dependent on calibration gas
accuracy and time since last calibration
Zero Drift:
Less than 0.1% of full-scale per month,
non-cumulative
Span Drift:
Application dependent, but generally less
than 3% per month
Linearity:
+/- 0.5% of full-scale
Repeatability:
+/- 1% of full-scale
Temperature Range:
0°F to +120°F (-18°C to +49°C)
Humidity Range:
5% to 100% condensing
Dimensions:
7.7” high x 6.7” wide x 3.8” deep
Weight:
3 lbs
Wiring Connections:
3-conductor, shielded, stranded,
≥ 20 AWG cable (General Cable C2525A or equivalent)
up to 1500 ft.
Terminal Block Plug (Field Wiring):
26-12 AWG,
torque 4.5 lbs-in.
Enclosure:
Injection-molded, NEMA 3RX washdown-
duty, polycarbonate sensor housing with hinged lid
and captive screw. For non-classified areas.
Optional 18 GA, NEMA 3RX washdown-duty stainless
steel enclosure with hinged lid and captive screw. For
non-classified areas.
Certification:
ETL Listed:
Conforms to UL 61010-1
Certified to CSA C22.2 No. 61010-1