2.4

  

Installation / Operation

2.4.1

  

Installing and Operating the 2116-21#-1014 Downconverter (With connector option SN)

1.) Connect  a -20 dBm to -0 dBm signal to RF INPUT, J1 (Figure 2.1).
2.) Connect the RF OUT, J101, to the external equipment.
3.) Connect 100- 240 ±10% VAC, 47 - 63 Hz to AC connector on the back panel.
4.) Be sure DS1 (green, DC Power) is on and DS2 (red, Alarm) is off (Figure 2.2).
5.) Select either INT (for internal 10MHz ref), AUTO (for internal 10MHz ref UNLESS a 
     external 10MHz, 0 dBm signal is connected to J2), or EXT (for external 10MHz, 
     0 dBm ref that is inserted at J2) on rear panel switch S1 (Figure 2.1).
6.) If EXT is selected or AUTO is selected and there is a 10MHz, 0 dBm signal at J2, 
      check that DS3 (yellow, EXT 10MHZ) is on (Figure 2.2).
7.) Check that a 10MHz, 0 dBm ±3 dB signal is present at the 10MHZ REF OUTPUT (J8) (Figure 2.1).
8.) 

AC Fuse - 

The fuse is a 5 mm X 20 mm, 2 amp slow blow (Type T) and is inserted in the far slot 

     in the drawer below the AC input as shown in Figure 2.3. There is a spare fuse in the near slot. 
     If a fuse continues to open, the power supply is most likely defective.

AC Fuse - 2 amp slow blow (Type T), 
5 mm X 20 mm

FUSE DRAWER

SPARE FUSE

~

INPUT

100-240± 10%VAC

47-63 Hz

2A MAX

FUSE

TYPE T 2A GDC

250 VOLT

FOR 100 - 240 V~

~

FIGURE 2.3    Fuse Location and Spare Fuse

2116-21#-1014 Manual

 

  Page 8

    

       03/18/21

Содержание 2116-21-1014

Страница 1: ...eproduced or duplicated without the prior permission of Cross Technologies Inc for purposes of operating the equipment When ordering parts from Cross Technologies Inc be sure to include the equipment...

Страница 2: ...ogies Inc s obligation under this warranty is limited to repairing or at Cross Technologies Inc s option replacing parts subassemblies or entire assemblies Cross Technologies Inc shall not be liable f...

Страница 3: ...he external reference is used and in the AUTO position the internal reference is used unless a 3dBm 3dB 10MHz signal is connected to the external reference input The 2116 21 1014 is powered by a 100 2...

Страница 4: ...high X 14 0 deep Size Other 100 240 10 VAC 47 63 Hz 25 watts maximum Power DB9 NO or NC contact closure on Alarm Alarm Connector BNC femal 50 75 10 MHz Connectors N Type female 50 BNC female 50 RF L B...

Страница 5: ...RF 50 N Type L BAND SN 50 SMA RF 50 SMA L BAND SS Available Connector Options 50 SMA RF 75 F Type L BAND SF 50 SMA RF 75 BNC L BAND S7 50 N Type RF 50 N Type L BAND NN 50 N Type RF 75 F Type L BAND NF...

Страница 6: ...s The 2116 21 1014 can be secured to a rack using the 4 holes on the front panel Figure 2 0 shows how the 2116 21 1014 is assembled 1 PS IN PIN 1 12 VDC PIN 2 5 VDC PIN 3 5 VDC PIN 4 GROUND PIN 5 GROU...

Страница 7: ...UTPUT BNC female 75 Outputs currently selected 10MHz ref signal 3 dBm 3 dB FIGURE 2 1 2116 21 1014 Rear Panel I O s shown with connector option SN TABLE 2 1 J11 Pinouts DB9 TABLE 2 1 J11 Pinouts DB9 P...

Страница 8: ...ted at J2 on rear panel switch S1 Figure 2 1 6 If EXT is selected or AUTO is selected and there is a 10MHz 0 dBm signal at J2 check that DS3 yellow EXT 10MHZ is on Figure 2 2 7 Check that a 10MHz 0 dB...

Страница 9: ...equipment is not compromised Additional space between units may be required E Circuit Overloading Consideration should be given to the connection of the equipment to the supply circuit and the effect...

Страница 10: ...S INC 6170 Shiloh Road Alpharetta Georgia 30005 770 886 8005 FAX 770 886 7964 Toll Free 888 900 5588 WEB www crosstechnologies com E MAIL info crosstechnologies com Printed in USA 2116 21 1014 Manual...

Отзывы: