© 2022 All rights reserved. Data subject to change without notice.
11
Rev. C | 2022.02
FCS - Operation Manual
Relays (internal)
4 internal SPDT dry contact relays, rated 5A @ 240 VAC
Audible Alarm
- Standard door mounted buzzer, rated 90 dB @ 30 cm (1 ft)
- Optional door mounted water tight buzzer (Option -WA), rated
85 dB @ 60.96 cm / 2 ft
Top Mounted Strobe
24 V, 114 mm H x 76 mm dia / 4.5 in H x 3 in diameter
(Option -L)
ELECTRICAL
Power Requirement
90 - 240 VAC, 50 - 60 Hz, 75 W Line Voltage
Current Draw
Line Voltage (110 VAC)
Line Voltage (220 VAC)
Approximately 1.0 Amp
Approximately 0.5 Amp
Power Distribution
Total power available to Remote and Peripheral Devices and
Options 65 W @ 24 VDC
Wiring
- VAC (line voltage) three-conductor (Line, Neutral, Ground)
shielded minimum 18 AWG stranded within conduit
- LAN Modbus
®
RTU RS-485 4-conductor, 16 AWG, stranded
shielded
- WAN (output to BAS): 4-conductor, 16 AWG, stranded shielded
Modbus
®
RTU (version 1.1b3) RS-485 or
BACnet® MS/TP (version 1 rev 14) RS-485
LAN / WAN
Communication:
Modbus
®
RTU over
RS-485
LAN Baud rate: 19,200 (default, configurable)
WAN Baud rate: 19,200 (default, configurable)
WAN Modbus ID: 100 (factory default)
Data bits: 8
Stop bits: 1
Parity: none
WAN Communication:
BACnet
®
MS/TP
Baud rate: 76,800 (default, configurable)
Base Address: 270 (factory default)
MAC Address: 100 (factory default)
Instance ID: 270100 (the Base Address x 1000 + the MAC
Address)
Data Bits: 8
Stop Bits: 1
Parity: none
Fuses
Automatic resetting thermal
ENVIRONMENTAL
Operating Temperature
-20°C to 40°C (-4°F to 104°F)
Operating Humidity
15 - 90% RH non-condensing