![CPS FA1000 SERIES Скачать руководство пользователя страница 19](http://html.mh-extra.com/html/cps/fa1000-series/fa1000-series_operation-manual_2675201019.webp)
To enter into the units
MAINTENANCE
mode, hold down
Key 2
and then turn power switch on.
MAINTENANCE VIA KEYPAD OPERATION
The LCD will now read:
FILTER CHANGE: Push the
CHANGE FILTER
key.
Go to page 21 for detailed FILTER CHANGE instructions.
MAINTENANCE MODE
Key 2
SCALE CHECK:
Push the
SCALE
key. The LCD will now read:
Push the
YES
key to proceed with accuracy check.
Pushing the
NO
key will skp the accuracy check and
proceed to the
Re-Zeroing
function.
CHECK SCALE ACCURACY?
YES >
NO >
PLACE CALIBRATION
WEIGHT ON SCALE
DOES READING MATCH
CALIBRATION WEIGHT?
0 g
YES >
NO >
The LCD will now read:
Place Calibration Weight on tank.
Push the
YES
key if weight reading matches Calibration weight.
(Push the
NO
key if weight does not match. See service manual
for complete scale re-calibration procedure. A 25LB or 15KG
Class F weight will be required for scale re-calibration).
The LCD will now read:
Push the
YES
key to proceed with
RE-ZERO
scale.
Note: Pushing the NO key will prompt the unit to
reboot and Go to MAIN MENU.
LCD will now read:
Lift the Tank off the scale.
Then Push
NEXT
key.
SCALE RE-ZERO
FUNCTION?
CURRENT SCALE
WEIGHT
0 g
YES >
NO >
REMOVE TANK
FROM SCALE
PLATFORM
PUSH [NEXT]
NEXT >
QUIT >
19
MAINTENANCE
SELECT MODE
CHANGE FILTER >
SCALE >
PT ZERO >
CHANGE REF >
QUIT >
MAINTENANCE
Содержание FA1000 SERIES
Страница 24: ...www cpsproducts com 73 088 REV B ...