Scanning Electron Microscope
EM-30PLUS Operating Manual
23
Gun Align
: Align the position of electron beam generated from e-gun.
Gun Align must be performed when replacing Filament or changing acceleration
voltage.
Find out the position where image gets brightest by dragging X, Y axis or adjusting
with mouse wheel.
- After e-gun alignment of hardware, be sure to carry out gun alignment in NS.
- When using with Gun align biased excessively, it can be hard on equipment.
Center X, Y: If you press X button, the position of X Gun Align moves to the center. If
you press Y button, the position of Y Gun Align moves to the center.
(As the position of center does not necessarily mean the optimum state of e-gun, carry
out Fine Adjustment as the brightest position after moving it to the right center.)
2.6.2 Magnification
: You can select or input the magnification you want while observing specimens.
Magnification: Select the magnification you want by using mouse wheel or inputting
figures.
(Range of magnification setting: x15~x150,000 times)
The minimum and maximum magnification refers to the value that does not drop any
more, which varies depending on the Working Distance(WD).
Fixed Magnification button(X200, 500, 1000, 2000, 5000, 10000): Select the
magnification you want by pressing the button.