11.0 COVVI Go App
For Fast Open/Slow Close the following parameters can be established:
Fast Open / Slow Close Time -
A signal sent within this time is interpreted as an open signal and a signal sent
after this time is interpreted as a close signal.
Long Hold Open Time -
This is the time an open signal must be maintained once the hand is fully opened for
the Long Hold Open grip switch trigger to work.
Input A or B Speed -
The maximum speed of the fingers and thumb can be reduced. It can be set between 50%
and 100%. As the one input is used for both opening and closing the hand, if you reduce the percentage the
speed of opening and closing the hand will be reduced. You cannot reduce the speed for just opening or just
closing the hand. This is a feature designed to aid new users who are still adapting to using a prosthetic device.
For Single Site Alternating the following parameters can be defined:
Direction Change Time -
A signal sent after this time will change the direction of the hand and a signal sent
within this time will keep the direction of the hand the same.
Long Hold Open Time -
This is the time an open signal must be maintained once the hand is fully opened for
the Long Hold Open grip switch trigger to work.
Input A or B Speed -
The maximum speed of the fingers and thumb can be reduced. It can be set between 50%
and 100%. As the one input is used for both opening and closing the hand, if you reduce the percentage the
speed of opening and closing the hand will be reduced. You cannot reduce the speed for just opening or just
closing the hand. This is a feature designed to aid new users who are still adapting to using a prosthetic device.
Содержание HAND
Страница 2: ...Anything Is Possible...