6.0 Fitting the COVVI Hand
6.3 System Notifications
When Auto Grip is enabled, if the hand senses an object is slipping it reapplies a small continuous force to
grasp the object. Auto Grip works in power grip, tripod grip, precision open and precision closed.
Auto Relax moves the hand to a relaxed position after a period of inactivity. A user input signal will cause the
hand to revert to its previous grip and the icon will be removed.
A short press to the dorsal button (under two seconds) puts the hand into Standby Mode where user inputs
from the electrodes have no effect on the hand. Pressing the dorsal button again will exit Standby Mode and
the icon will be removed.
The CAN bus icon is displayed whenever the hand is connected to a CAN enabled device, such as the COAPT
Gen2 system. NOTE: this icon is only displayed once Bluetooth has turned off (the two icons share the same
location on the display).
The warning icon shows that the hand has detected a fault. The screen will also display text to identify what
the fault is, for example, 1 of 1 Fault Thumb Sensor.
Fault detected
Connected to
CAN bus
Standby Mode is activated
Auto Relax is enabled
Auto Grip is enabled