SatSeconds™ Alarm Management Feature
Operator’s Manual
87
10.6.2
Second SpO
2
Event
Consider the second event. Suppose the SatSeconds alarm limit is still set to
25. The patient’s SpO
2
drops to 84% and the duration of the event is 15
seconds before saturation again exceeds the lower alarm threshold of 85%.
Because the SatSeconds alarm limit is set to 25 and the actual number of Sat-
Seconds equals 15, there is no audible alarm.
Figure 10-4.
Second SpO
2
Event: No SatSeconds Alarm
1% drop below the lower alarm limit threshold
x15 second duration below the lower threshold
15 SatSeconds
; no alarm