
Theory of Operations
86
Operator’s Manual
10.6.1
First SpO
2
Event
Consider the first event. Suppose the SatSeconds alarm limit is set to 25. The
patient’s SpO
2
drops to 79% and the duration of the event is two (2) seconds
before saturation again exceeds the lower alarm threshold of 85%.
Because the SatSeconds alarm limit is set to 25 and the actual number of Sat-
Seconds equals 12, there is no audible alarm.
Figure 10-3.
First SpO
2
Event: No SatSeconds Alarm
6% drop below the lower alarm limit threshold
x 2 second duration below the lower threshold
12 SatSeconds
; no alarm
Содержание Nellcor
Страница 1: ...Operator s Manual Nellcor Bedside SpO2 Patient Monitoring System TM ...
Страница 8: ...viii ...
Страница 10: ...Page Left Intentionally Blank x ...
Страница 18: ...Introduction 8 Operator s Manual Page Left Intentionally Blank ...
Страница 100: ...Page Left Intentionally Blank 90 Operator s Manual Theory of Operations ...
Страница 119: ......