User Interface
Operator’s Manual
27
To power on the monitoring system:
1.
Press the
Power On/Off Button
for more than one (1) second.
2.
Ensure the software version, the SpO
2
alarm indicator, and the pulse rate alarm
indicator light for approximately two (2) seconds.
Figure 4-1.
Sample Initial Screen
3.
Ensure the
POST pass
tone sounds when POST completes.
Note:
Do not use the monitoring system should a repeating, high-pitched alarm tone occur at
power on. Instead, please contact Technical Services or a qualified service technician.
4.3.2
Turning off the Monitoring System
After using the monitoring system, turn it off safely.
To turn off the monitoring system:
1.
Press the Power On/Off button on the right of the device for approximately one second.
2.
Observe the message
System is shutting down
on the screen.
Note:
Press the Power On/Off button for at least 15 seconds to turn off the monitoring
system after any situation involving continuous resets or a system lock.
Содержание Nellcor
Страница 1: ...Operator s Manual Nellcor Bedside SpO2 Patient Monitoring System TM ...
Страница 8: ...viii ...
Страница 10: ...Page Left Intentionally Blank x ...
Страница 18: ...Introduction 8 Operator s Manual Page Left Intentionally Blank ...
Страница 100: ...Page Left Intentionally Blank 90 Operator s Manual Theory of Operations ...