Error Conditions
Operator’s Manual
75
Blank screen
Ensure buttons light. If not, press the Power On/Off button to turn on.
Check if AC indicator lights or blinks.
Use the same AC power source with other equipment to check for power.
If the error continues, contact Technical Services or a qualified service tech-
nician.
Screen does not function prop-
erly and the power-on beep
tones do not sound
Do not use the monitoring system; contact a qualified service technician or
Covidien Technical Services.
No sound generation
Verify setting point of volume is not 0 or 1.
Verify alarm setup is not set to audible alarm paused.
If the error continues, contact Technical Services or a qualified service tech-
nician.
Abnormally shut down last
time message
Check any temporary settings such as alarm limits, response mode, and
patient mode, since resets invoke factory or institutional default settings.
Press the Power On/Off button to reset system power.
If the error continues, contact Technical Services or a qualified service tech-
nician.
Date and Time incorrect
Set the time in the Options Menu.
Check if date setting format follows locale.
If the system displays wrong date and time even after power reset, it means
internal battery for power backup is dead.
If the error continues, contact Technical Services or a qualified service tech-
nician.
System consumes battery
power even with AC power
connection
Ensure proper connection between power cord and wall socket.
Check if AC indicator lights or blinks.
Use the same AC power source with other equipment to check for power.
Replace the power cord.
If the error continues, contact Technical Services or a qualified service tech-
nician.
Low Battery / Critically Low-
Battery condition
Connect the system to AC power until the internal battery is fully charged.
Ensure the system power cord is connected to the wall socket properly.
Check if AC indicator lights or blinks.
Use the same AC power source with other equipment to check for power.
Check the date of manufacture (DOM) of battery.
If the error continues, contact Technical Services or a qualified service tech-
nician.
Table 8-1.
Common Problems and Resolutions (Continued)
Problem
Resolution
Содержание Nellcor
Страница 1: ...Operator s Manual Nellcor Bedside SpO2 Patient Monitoring System TM ...
Страница 8: ...viii ...
Страница 10: ...Page Left Intentionally Blank x ...
Страница 18: ...Introduction 8 Operator s Manual Page Left Intentionally Blank ...
Страница 100: ...Page Left Intentionally Blank 90 Operator s Manual Theory of Operations ...
Страница 119: ......